NEMA NU 4-2008 Performance Measurements of Two Commercial Small-Animal PET Scanners: ClearPET and rPET-1
- 9 November 2010
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 58 (1) , 58-65
Abstract
In this work, we compare two commercial positron emission tomography (PET) scanners installed at CIEMAT (Madrid, Spain): the ClearPET and the rPET-1. These systems have significant geometrical differences, such as the axial field of view (110 mm on ClearPET versus 45.6 mm on rPET-1), the configuration of the detectors (whole ring on ClearPET versus one pair of planar blocks on rPET-1) and the use of an axial shift between ClearPET detector modules. We used an assessment procedure that fulfilled the recommendations of the National Electrical Manufacturers Association (NEMA) NU 4-2008 standard. The methodology includes studies of spatial resolution, sensitivity, scatter fraction, count losses and image quality. Our experiments showed a central spatial resolution of 1.5 mm (transaxial), 3.2 mm (axial) for the ClearPET and 1.5 mm (transaxial), 1.6 mm (axial) for the rPET-1, with a small variation across the transverse axis on both scanners (~1 mm). The absolute sensitivity at the centre of the field of view was 4.7% for the ClearPET and 1.0% for the rPET-1. The peak noise equivalent counting rate for the mouse-sized phantom was 73.4 kcps reached at 0.51 MBq/mL on the ClearPET and 29.2 kcps at 1.35 MBq/mL on the rPET-1. The recovery coefficients measured using the image quality phantom ranged from 0.11 to 0.89 on the ClearPET and from 0.14 to 0.81 on the rPET-1. The overall performance shows that both the ClearPET and the rPET-1 systems are very suitable for preclinical research and imaging of small animals.Keywords
This publication has 18 references indexed in Scilit:
- Performance Evaluation of a PEM Scanner Using the NEMA NU 4—2008 Small Animal PET StandardsIEEE Transactions on Nuclear Science, 2010
- Design and performance evaluation of a coplanar multimodality scanner for rodent imagingPhysics in Medicine & Biology, 2009
- Performance Evaluation of the Inveon Dedicated PET Preclinical Tomograph Based on the NEMA NU-4 StandardsJournal of Nuclear Medicine, 2009
- Positron Emission Tomography Imaging as a Key Enabling Technology in Drug Development2006
- The data acquisition system of ClearPET neuro - a small animal PET scannerIEEE Transactions on Nuclear Science, 2006
- rPET Detectors Design and Data Processing2006
- Small animal PET: aspects of performance assessmentEuropean Journal of Nuclear Medicine and Molecular Imaging, 2004
- Design optimization of the PMT-ClearPET prototypes based on simulation studies with GEANT3IEEE Transactions on Nuclear Science, 2003
- Accelerated image reconstruction using ordered subsets of projection dataIEEE Transactions on Medical Imaging, 1994
- Analytic 3D image reconstruction using all detected eventsIEEE Transactions on Nuclear Science, 1989