rPET Detectors Design and Data Processing
Authors:
- E. LageELE. Lage
- L. RiconLRL. Ricon
- M. AbellaMAM. Abella
- E. VicenteEVE. Vicente
- M. DescoMDM. Desco
- 21 March 2006
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- Vol. 5 (10823654) , 2885-2889
Abstract
Small animal PET systems based on rotating planar detectors posses some interesting advantages for high sensitivity, high resolution imaging. We have designed the rPET detectors based on MLS crystals assembled on a 30times30 matrix optically coupled to a flat-panel PS-PMT. Weighted position readout circuits pre-process the 64 signals from the 8times8 anodes matrix, which are digitized using a charge-integrating converter. The amplification electronics, including the trigger output for coincidence detection, and the high voltage supply are integrated in a three PCBs stack that forms the base attached to the back of the PMT. The whole assembly is enclosed in a light tight, lead (Pb) shielded aluminum box. The detectors are mounted on a rotating gantry with more than 180 degrees rotation span. The digitized events are screened and histogramed, and a modified center of gravity algorithm removes from the position calculation those signals with poor signal to noise ratio. Apparent mean crystal size on the 511 keV field-flood images is 0.6 mm, mean peak-to-valley ratio is better than 8, and intrinsic resolution is 1.5 mm at the central row, with the energy window wide open. Sensitivity (CPS) for a pair of these detectors set in coincidence at 160 mm distance is 1%Keywords
This publication has 24 references indexed in Scilit:
- Performance characteristics of a second generation micro crystal element (MiCE2) detector2005
- Investigation of signal readout methods for the Hamamatsu R8500 flat panel PSPMT2004
- Performance of a PET Detector With a 256ch Flat Panel PS-PMTIEEE Transactions on Nuclear Science, 2004
- Compact readout electronics for position sensitive photomultiplier tubes2003
- Readout electronics for multianode photomultiplier tubes with pad matrix anode layout2003
- High spatial resolution scintillation detector based on the H8500 photomultiplier2003
- Analog readout system with charge division type output2002
- Depth identification accuracy of a three layer phoswich PET detector moduleIEEE Transactions on Nuclear Science, 1999
- Characteristics of a pair of small field-of-view LSO scintillation camerasIEEE Transactions on Nuclear Science, 1996
- Cerium-doped lutetium oxyorthosilicate: a fast, efficient new scintillatorIEEE Transactions on Nuclear Science, 1992