Numerical, Hydraulic, and Hemolytic Evaluation of an Intravascular Axial Flow Blood Pump to Mechanically Support Fontan Patients
- 14 September 2010
- journal article
- research article
- Published by Springer Nature in Annals of Biomedical Engineering
- Vol. 39 (1) , 324-336
Abstract
Currently available mechanical circulatory support systems are limited for adolescent and adult patients with a Fontan physiology. To address this growing need, we are developing a collapsible, percutaneously-inserted, axial flow blood pump to support the cavopulmonary circulation in Fontan patients. During the first phase of development, the design and experimental evaluation of an axial flow blood pump was performed. We completed numerical modeling of the pump using computational fluid dynamics analysis, hydraulic testing of a plastic pump prototype, and blood bag experiments (n = 7) to measure the levels of hemolysis produced by the pump. Statistical analyses using regression were performed. The prototype with a 4-bladed impeller generated a pressure rise of 2–30 mmHg with a flow rate of 0.5–4 L/min for 3000–6000 RPM. A comparison of the experimental performance data to the numerical predictions demonstrated an excellent agreement with a maximum deviation being less than 6%. A linear increase in the plasma-free hemoglobin (pfHb) levels during the 6-h experiments was found, as desired. The maximum pfHb level was measured to be 21 mg/dL, and the average normalized index of hemolysis was determined to be 0.0097 g/100 L for all experiments. The hydraulic performance of the prototype and level of hemolysis are indicative of significant progress in the design of this blood pump. These results support the continued development of this intravascular pump as a bridge‐to‐transplant, bridge‐to‐recovery, bridge-to-hemodynamic stability, or bridge-to-surgical reconstruction for Fontan patients.Keywords
This publication has 22 references indexed in Scilit:
- Interaction of an Idealized Cavopulmonary Circulation With Mechanical Circulatory Assist Using an Intravascular Rotary Blood PumpArtificial Organs, 2010
- Hydraulic Testing of Intravascular Axial Flow Blood Pump Designs With a Protective Cage of Filaments for Mechanical Cavopulmonary AssistASAIO Journal, 2010
- Intravascular Mechanical Cavopulmonary Assistance for Patients With Failing Fontan PhysiologyArtificial Organs, 2009
- An Artificial Right Ventricle for Failing Fontan: In Vitro and Computational StudyThe Annals of Thoracic Surgery, 2009
- Long-Term Survival, Modes of Death, and Predictors of Mortality in Patients With Fontan SurgeryCirculation, 2008
- Factors influencing early and late outcome following the Fontan procedure in the current era. The ‘Two Commandments’?☆European Journal of Cardio-Thoracic Surgery, 2007
- Fluid Mechanic Assessment of the Total Cavopulmonary Connection using Magnetic Resonance Phase Velocity Mapping and Digital Particle Image VelocimetryAnnals of Biomedical Engineering, 2000
- Plasma Hemoglobin Measurement Techniques for the In Vitro Evaluation of Blood Damage Caused by Medical DevicesArtificial Organs, 1997
- Use of computational fluid dynamics in the design of surgical procedures: Application to the study of competitive flows in cavopulmonary connectionsThe Journal of Thoracic and Cardiovascular Surgery, 1996
- Three‐Dimensional Numerical Prediction of Stress Loading of Blood Particles in a Centrifugal PumpArtificial Organs, 1995