Phantom Validation of Spectral Detector Computed Tomography–Derived Virtual Monoenergetic, Virtual Noncontrast, and Iodine Quantification Images
- 1 November 2018
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Computer Assisted Tomography
- Vol. 42 (6) , 959-964
Abstract
Spectral detector computed tomography (SDCT) is a new CT technology that uses a dual-layer detector to perform energy separation. We aim to assess 3 clinical concepts using a phantom model: noise profile across the virtual monoenergetic (VME) spectrum, accuracy of iodine quantification, and virtual noncontrast (VNC) reconstructions' ability to remove iodine contribution to attenuation. Six vials containing varying concentrations of iodinated contrast (0–6 mg/mL) diluted in water were placed in a water bath and scanned on an SDCT scanner. Virtual monoenergetic (40–200 keV at 10-keV increments), iodine–no-water, and VNC reconstructions were created. Attenuation (in Hounsfield units [HU]), VME noise at each energy level, CT-derived iodine concentration, and VNC attenuation were recorded. Virtual monoenergetic noise was improved at all energies compared with conventional images (conventional, 9.8–11.2; VME, 7.5–9.5). Noise profile showed a slightly higher image noise at 40 keV, but was otherwise relatively flat across the energy spectrum. On iodine–no-water reconstructions, measured varied from actual iodine concentration by ±0.1 mg/mL (SD, 0.16–0.36). Virtual noncontrast attenuation was within 5 HU of water attenuation at all iodine concentrations. Reconstructions of SDCT show lower VME image noise, accurate iodine quantification, and VNC attenuation values within 5 HU of expected in a phantom model.Keywords
This publication has 24 references indexed in Scilit:
- Spectral detector CT-derived virtual non-contrast images: comparison of attenuation values with unenhanced CTAbdominal Radiology, 2017
- Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical ApplicationsRadiology, 2015
- Maximizing Iodine Contrast-to-Noise Ratios in Abdominal CT Imaging through Use of Energy Domain Noise Reduction and Virtual Monoenergetic Dual-Energy CTRadiology, 2015
- Dual-Energy MDCT for Imaging the Renal MassAmerican Journal of Roentgenology, 2015
- Dual-energy CT of the abdomenAbdominal Radiology, 2013
- Virtual Monochromatic Reconstruction of Dual-Energy Unenhanced Head CT at 65–75 keV Maximizes Image Quality Compared with Conventional Polychromatic CTRadiology, 2013
- Virtual Monochromatic Spectral Imaging for the Evaluation of Hypovascular Hepatic MetastasesInvestigative Radiology, 2012
- Characterization of Adrenal Nodules With Dual-Energy CT: Can Virtual Unenhanced Attenuation Values Replace True Unenhanced Attenuation Values?American Journal of Roentgenology, 2012
- Dual-Energy CT in Patients Suspected of Having Renal Masses: Can Virtual Nonenhanced Images Replace True Nonenhanced Images?Radiology, 2009
- Energy-selective reconstructions in X-ray computerised tomographyPhysics in Medicine & Biology, 1976