A Laplace–Hamming Binarization Approach for Second‐Generation HR‐pQCT Rescues Fine Feature Segmentation
Authors:
Open Access
- 13 May 2023
- journal article
- research article
- Published by Oxford University Press (OUP) in Journal of Bone and Mineral Research
- Vol. 38 (7) , 1006-1014
Abstract
Although second‐generation high‐resolution peripheral quantitative computed tomography (XCTII) provides the highest‐resolution in vivo bone microstructure assessment, the manufacturer's standard image processing protocol omits fine features in both trabecular and cortical compartments. To optimize fine structure segmentation, we implemented a binarization approach based on a Laplace–Hamming (LH) segmentation and documented the reproducibility and accuracy of XCTII structure segmentation using both the standard Gaussian‐based binarization and the proposed LH segmentation approach. To evaluate reproducibility, 20 volunteers (9 women, 11 men; aged 23–75 years) were recruited, and three repeat scans of the radii and tibias were acquired using the manufacturer's standard in vivo protocol. To evaluate accuracy, cadaveric structure phantoms (14 radii, 6 tibias) were scanned on XCTII using the same standard in vivo protocol and on μCT at 24.5 μm resolution. XCTII images were analyzed twice—first, with the manufacturer's standard patient evaluation protocol and, second, with the proposed LH segmentation approach. The LH approach rescued fine features evident in the grayscale images but omitted or overrepresented (thickened) by the standard approach. The LH approach significantly reduced error in trabecular volume fraction (BV/TV) and thickness (Tb.Th) compared with the standard approach; however, higher error was introduced for trabecular separation (Tb.Sp). The LH approach improved the correlation between XCTII and μCT for cortical porosity (Ct.Po) and significantly reduced error in cortical pore diameter (Ct.Po.Dm) compared with the standard approach. The LH approach resulted in improved precision compared with the standard approach for BV/TV, Tb.Th, Ct.Po, and Ct.Po.Dm at the radius and for Ct.Po at the tibia. Our results suggest that the proposed LH approach produces substantially improved binary masks, reduces proportional bias, and provides greater accuracy and reproducibility in important outcome metrics, all due to more accurate segmentation of the fine features in both trabecular and cortical compartments. © 2023 The Authors. Journal of Bone and Mineral Research published by Wiley Periodicals LLC on behalf of American Society for Bone and Mineral Research (ASBMR).Keywords
Funding Information
- National Institute of Arthritis and Musculoskeletal and Skin Diseases (P30 AR075055, R01 AR060700)
This publication has 21 references indexed in Scilit:
- Multicenter precision of cortical and trabecular bone quality measures assessed by high-resolution peripheral quantitative computed tomographyJournal of Bone and Mineral Research, 2012
- Reproducibility of direct quantitative measures of cortical bone microarchitecture of the distal radius and tibia by HR-pQCTBone, 2010
- Microarchitectural deterioration of cortical and trabecular bone: Differing effects of denosumab and alendronateJournal of Bone and Mineral Research, 2010
- Early impairment of trabecular microarchitecture assessed with HR-pQCT in patients with stage II-IV chronic kidney diseaseJournal of Bone and Mineral Research, 2010
- Bone micro-architecture and determinants of strength in the radius and tibia: age-related changes in a population-based study of normal adults measured with high-resolution pQCTOsteoporosis International, 2009
- Bone strength at the distal radius can be estimated from high-resolution peripheral quantitative computed tomography and the finite element methodBone, 2008
- In Vivo Determination of Bone Structure in Postmenopausal Women: A Comparison of HR-pQCT and High-Field MR ImagingJournal of Bone and Mineral Research, 2008
- Accuracy of high-resolution peripheral quantitative computed tomography for measurement of bone qualityMedical Engineering & Physics, 2007
- A Local Adaptive Threshold Strategy for High Resolution Peripheral Quantitative Computed Tomography of Trabecular BoneAnnals of Biomedical Engineering, 2007