Design of microfluidic channels to prevent negative filtration in implantable hemofiltration devices
- 1 November 2021
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE) in Annual International Conference of the IEEE Engineering in Medicine and Biology Society
- Vol. 2021, 5051-5054
Abstract
In order to improve the quality of life of dialysis patients, our group have been developing an implantable hemofiltration device (IHFD) composed of multiple layers of dialysis membranes and microfluidic channels. To improve the hemodialysis performance of IHFD, preventing the negative filtration, which is caused by the oncotic pressure of blood, is mandatory. In this study, we fabricated IHFDs with five different microchannel designs and experimentally investigated the performance of each device in in vitro experiment. In addition, the successful IHFD was further evaluated by ex vivo experiments with a beagle dog. The experiments verified the effectiveness of the microchannel design, which will be used for the IHFD for in vivo experiments with pigs in the future.Keywords
This publication has 6 references indexed in Scilit:
- Radiation sterilization of tissue allografts: A reviewWorld Journal of Radiology, 2016
- Water-Permeable Dialysis Membranes for Multi-Layered Microdialysis SystemFrontiers in Bioengineering and Biotechnology, 2015
- Associations of Pretransplant Serum Albumin with Post-Transplant Outcomes in Kidney Transplant RecipientsAmerican Journal of Transplantation, 2011
- The use of 25% human serum albumin: outcome and efficacy in raising serum albumin and systemic blood pressure in critically ill dogs and catsJournal of Veterinary Emergency and Critical Care, 2005
- Backfiltration: A Controversial Issue in Modern DialysisThe International Journal of Artificial Organs, 1988
- A canine ex vivo series shunt for evaluating thrombus deposition on polymer surfacesJournal of Biomedical Materials Research, 1984