Power Supply Topologies for Biphasic Stimulation in Inductively Powered Implants
- 27 July 2005
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 2743
Abstract
Biphasic stimulation is commonly used for electrical stimulation of tissue in many implants. A dual voltage scheme is often used for the operation of the stimulus circuitry. Due to the advantages of wireless power transmission, an inductive link is used to recover the transmitted power and generate these voltages. This paper identifies the problem of additional power dissipation in the implant when the traditional rectifier topology with one diode is used. A dual diode topology is proposed which reduces the power dissipated in the implant by half. A detailed analysis for both topologies is done through derivation of expressions for the power losses on the primary and secondary coils. A comparison of single diode and dual diode topologies is presented under different conditions and inferences are drawn that can be used as design guidelines.Keywords
This publication has 6 references indexed in Scilit:
- An efficient inductive power link design for retinal prosthesis2004
- A matched biphasic microstimulator for an implantable retinal prosthetic device2004
- A smart bi-directional telemetry unit for retinal prosthetic device2003
- A single-channel implantable microstimulator for functional neuromuscular stimulationIEEE Transactions on Biomedical Engineering, 1997
- An advanced multiple channel cochlear implantIEEE Transactions on Biomedical Engineering, 1989
- Design of radio-frequency powered coils for implant instrumentsMedical & Biological Engineering & Computing, 1977