A handheld luminometer with sub-attomole limit of detection for distributed applications in global health
Authors:
- Paul LebelPLPaul Lebel
- Susanna ElledgeSESusanna Elledge
- Diane M. WienerDWDiane M. Wiener
- Ilakkiyan JeyakumarIJIlakkiyan Jeyakumar
- Maíra PhelpsMPMaíra Phelps
- Axel JacobsenAJAxel Jacobsen
- Emily HuynhEHEmily Huynh
- Chris CharltonCCChris Charlton
- Robert PuccinelliRPRobert Puccinelli
- Prasenjit MondalPMPrasenjit Mondal
- Senjuti SahaSSSenjuti Saha
- Cristina M. TatoCTCristina M. Tato
Open Access
- 21 February 2024
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS Global Public Health
- Vol. 4 (2) , e0002766
Abstract
Luminescence is ubiquitous in biology research and medicine. Conceptually simple, the detection of luminescence nonetheless faces technical challenges because relevant signals can exhibit exceptionally low radiant power densities. Although low light detection is well-established in centralized laboratory settings, the cost, size, and environmental requirements of high-performance benchtop luminometers are not compatible with geographically-distributed global health studies or resource-constrained settings. Here we present the design and application of a ~$700 US handheld, battery-powered luminometer with performance on par with high-end benchtop instruments. By pairing robust and inexpensive Silicon Photomultiplier (SiPM) sensors with a low-profile shutter system, our design compensates for sensor non-idealities and thermal drift, achieving a limit of detection of 1.6E-19 moles of firefly luciferase. Using these devices, we performed two pilot cross-sectional serology studies to assess sars-cov-2 antibody levels: a cohort in the United States, as well as a field study in Bangladesh. Results from both studies were consistent with previous work and demonstrate the device’s suitability for distributed applications in global health.Keywords
All Related Versions
Funding Information
- Chan Zuckerberg Biohub, San Francisco
This publication has 40 references indexed in Scilit:
- Using Split Luminescent Biosensors for SARS‐CoV‐2 Antibody Detection in Serum, Plasma, and Blood SamplesCurrent Protocols, 2022
- WHO International Standard for evaluation of the antibody response to COVID-19 vaccines: call for urgent action by the scientific communityThe Lancet Microbe, 2022
- Detection of SARS-CoV-2 IgG antibodies in dried blood spotsDiagnostic Microbiology and Infectious Disease, 2021
- IgG antibody response against nucleocapsid and spike protein post-SARS-CoV-2 infectionInfection, 2021
- Antibody response patterns in COVID‐19 patients with different levels of disease severity in JapanJournal of Medical Virology, 2021
- Integrating printed microfluidics with silicon photomultipliers for miniaturised and highly sensitive ATP bioluminescence detectionBiosensors and Bioelectronics, 2018
- Luminometer development in the last four decades: recollections of two entrepreneursLuminescence, 2012
- Characterisation studies of silicon photomultipliersNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2010
- Modeling and measurement of a whole-cell bioluminescent biosensor based on a single photon avalanche diodeBiosensors and Bioelectronics, 2008
- A portable silicon photodiode luminometerJournal of Bioluminescence and Chemiluminescence, 1987