Comparison of oscillometric, non-invasive and invasive arterial pressure monitoring in patients undergoing laparoscopic bariatric surgery – a secondary analysis of a prospective observational study
Open Access
- 28 March 2022
- journal article
- research article
- Published by Springer Nature in BMC Anesthesiology
- Vol. 22 (1) , 1-12
Abstract
Background: Oscillometric, non-invasive blood pressure measurement (NIBP) is the first choice of blood pressure monitoring in the majority of low and moderate risk surgeries. In patients with morbid obesity, however, it is subject to several limitations. The aim was to compare arterial pressure monitoring by NIBP and a non-invasive finger-cuff technology (Nexfin®) with the gold-standard invasive arterial pressure (IAP). Methods: In this secondary analysis of a prospective observational, single centre cohort study, systolic (SAP), diastolic (DAP) and mean arterial pressure (MAP) were measured at 16 defined perioperative time points including posture changes, fluid bolus administration and pneumoperitoneum (PP) in patients undergoing laparoscopic bariatric surgery. Absolute arterial pressures by NIBP, Nexfin® and IAP were compared using correlation and Bland Altman analyses. Interchangeability was defined by a mean difference ≤ 5 mmHg (SD ≤8 mmHg). Percentage error (PE) was calculated as an additional statistical estimate. For hemodynamic trending, concordance rates were analysed according to the Critchley criterion. Results: Sixty patients (mean body mass index of 49.2 kg/m2) were enrolled and data from 56 finally analysed. Pooled blood pressure values of all time points showed a significant positive correlation for both NIPB and Nexfin® versus IAP. Pooled PE for NIBP versus IAP was 37% (SAP), 35% (DAP) and 30% (MAP), for Nexfin versus IAP 23% (SAP), 26% (DAP) and 22% (MAP). Correlation of MAP was best and PE lowest before induction of anesthesia for NIBP versus IAP (r = 0.72; PE 24%) and after intraoperative fluid bolus administration for Nexfin® versus IAP (r = 0.88; PE: 17.2%). Concordance of MAP trending was 90% (SAP 85%, DAP 89%) for NIBP and 91% (SAP 90%, DAP 86%) for Nexfin®. MAP trending was best during intraoperative ATP positioning for NIBP (97%) and at induction of anesthesia for Nexfin® (97%). Conclusion: As compared with IAP, interchangeability of absolute pressure values could neither be shown for NIBP nor Nexfin®, however, NIBP showed poorer overall correlation and precision. Overall trending ability was generally high with Nexfin® surpassing NIBP. Nexfin® may likely render individualized decision-making in the management of different hemodynamic stresses during laparoscopic bariatric surgery, particularly where NIBP cannot be reliably established. Trial registration: The non-interventional, observational study was registered retrospectively at (NCT03184285) on June 12, 2017.Keywords
Funding Information
- Universitätsklinikum Schleswig-Holstein - Campus Kiel
This publication has 39 references indexed in Scilit:
- Perioperative management of the severely obese patient: a selective pathophysiological reviewCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 2012
- Noninvasive continuous hemodynamic monitoringJournal of Clinical Monitoring and Computing, 2012
- Noninvasive Continuous Arterial Blood Pressure Monitoring with Nexfin®Anesthesiology, 2012
- Noninvasive monitoring of blood pressure in the critically illCritical Care Medicine, 2012
- Investigation of the agreement of a continuous non-invasive arterial pressure device in comparison with invasive radial artery measurementBritish Journal of Anaesthesia, 2011
- Non-invasive continuous blood pressure monitoring of tachycardic episodes during interventional electrophysiologyEP Europace, 2010
- Effects of Bariatric Surgery on Mortality in Swedish Obese SubjectsThe New England journal of medicine, 2007
- Intraoperative Blood Pressure What Patterns Identify Patients at Risk for Postoperative Complications?Annals of Surgery, 1990
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- ERROR IN BLOOD-PRESSURE MEASUREMENT DUE TO INCORRECT CUFF SIZE IN OBESE PATIENTSThe Lancet, 1982