Polarization X-Ray Fluorescence Analysis by Milliwatt X-ray Tube
Open Access
- 15 September 2024
- journal article
- research article
- Published by Iron and Steel Institute of Japan in Tetsu-to-Hagane
- Vol. 110 (12) , 981-987
Abstract
X-ray spectra scattered to 90°were measured for various scatterers of 5-mm thick B4C, acrylic, titanium, iron, zirconium, and lead plates, and their combinations such as B4C(5 mm thick) on acrylic plate (1 or 5 mm thick). The degree of polarization was estimated from the observed X-ray spectra at 0°and 90°. Using the best scatterer as an X-ray polarizer for each samples, polarization X-ray fluorescence (XRF) spectra of various samples such as cabtyre cable, book cover made of synthetic-resin, and steel making slag, compared with non-polarization XRF spectra. Toxic elements in some samples were more sensitive for polarized XRF, while other samples were more sensitive for non-polarized XRF. The reasons for such contradicted experimental results were discussed. Most of the measurements were performed by 15 mW X-ray tube power for polarization XRF, and by 3 mW for non-polarized XRF. The meanings of lowest power XRF measurements in the literature were also discussed. This kind of low power experiments may open the possibility of easy way of experiments, and non-invasive XRF analysis for cultural heritage.Keywords
This publication has 11 references indexed in Scilit:
- X-Ray Spectroscopy for Chemical State Analysis2023
- Polarization and intensity of Compton scatteringX-Ray Spectrometry, 2021
- Polarized XRF Spectrometer with a 40-mW X-Ray TubeАналитика и контроль, 2018
- 3D-printed compact XRF spectrometerInternational Journal of PIXE, 2017
- Artificial peaks in energy dispersive X‐ray spectra: sum peaks, escape peaks, and diffraction peaksX-Ray Spectrometry, 2016
- Low-power total reflection X-ray fluorescence spectrometer using diffractometer guide railPowder Diffraction, 2014
- Effect of electrical charging on scanning electron microscopy-energy dispersive X-ray spectroscopy analysis of insulating materialsSpectrochimica Acta Part B: Atomic Spectroscopy, 2013
- Portable Total Reflection X-ray Fluorescence Spectrometer for Nanogram Cr Detection LimitAnalytical Chemistry, 2007
- X-ray fluorescence analysis at mg l−1level with an X-ray source powered by a dry batteryJournal of Analytical Atomic Spectrometry, 2004
- Site determination of oxygen inO by oxygenKα x-ray-emission spectroscopyPhysical Review B, 1990