Primary, scatter, and penetration characterizations of parallel-hole and pinhole collimators for I-123 SPECT
- 13 December 2019
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 64 (24) , 245001
Abstract
Multi-pinhole (MPH) collimators are known to provide better trade-off between sensitivity and resolution for preclinical, as well as for smaller regions in clinical SPECT imaging compared to conventional collimators. In addition to this geometric advantage, MPH plates typically offer better stopping power for penetration than the conventional collimators, which is especially relevant for I-123 imaging. The I-123 emits a series of high-energy (>300 keV, ~2.5% abundance) gamma photons in addition to the primary emission (159 keV, 83% abundance). Despite their low abundance, high-energy photons penetrate through a low-energy parallel-hole (LEHR) collimator much more readily than the 159 keV photons, resulting in large downscatter in the photopeak window. In this work, we investigate the primary, scatter, and penetration characteristics of a single pinhole collimator that is commonly used for I-123 thyroid imaging and our two MPH collimators designed for I-123 DaTscan imaging for Parkinson’s Disease, in comparison to three different parallel-hole collimators through a series of experiments and Monte Carlo simulations. The simulations of a point source and a digital human phantom with DaTscan activity distribution showed that our MPH collimators provide superior count performance in terms of high primary counts, low penetration, and low scatter counts compared to the parallel-hole and single pinhole collimators. For example, total scatter, multiple scatter, and collimator penetration events for the LEHR were 2.5, 7.6 and 14 times more than that of MPH within the 15% photopeak window. The total scatter fraction for LEHR was 56% where the largest contribution came from the high-energy scatter from the back compartments (31%). For the same energy window, the total scatter for MPH was 21% with only 1% scatter from the back compartments. We therefore anticipate that using MPH collimators, higher quality reconstructions can be obtained in a substantially shorter acquisition time for I-123 DaTscan and thyroid imaging.Keywords
Funding Information
- National Institute of Biomedical Imaging and Bioengineering (R01-EB022092)
This publication has 39 references indexed in Scilit:
- SNM Practice Guideline for Dopamine Transporter Imaging with 123I-Ioflupane SPECT 1.0Journal of Nuclear Medicine, 2011
- An investigation of inconsistent projections and artefacts in multi-pinhole SPECT with axially aligned pinholesPhysics in Medicine & Biology, 2011
- Review and current status of SPECT scatter correctionPhysics in Medicine & Biology, 2011
- Improved Quality of Pediatric 123I-MIBG Images with Medium-Energy CollimatorsJournal of Nuclear Medicine Technology, 2011
- Accurate Monte Carlo modelling of the back compartments of SPECT camerasPhysics in Medicine & Biology, 2010
- A fast cardiac gamma camera with dynamic SPECT capabilities: design, system validation and future potentialEuropean Journal of Nuclear Medicine and Molecular Imaging, 2010
- Performance evaluation of D-SPECT: a novel SPECT system for nuclear cardiologyPhysics in Medicine & Biology, 2009
- Small-Animal SPECT and SPECT/CT: Important Tools for Preclinical InvestigationJournal of Nuclear Medicine, 2008
- Monte Carlo simulation of energy spectra for123I imagingPhysics in Medicine & Biology, 2007
- Downscatter correction and choice of collimator in123I imagingPhysics in Medicine & Biology, 2006