CT dosimetry at the Australian Synchrotron for 25–100 keV photons and 35–160 mm-diameter biological specimens
- 15 February 2019
- journal article
- research article
- Published by International Union of Crystallography (IUCr) in Journal of Synchrotron Radiation
- Vol. 26 (2) , 517-527
Abstract
The dose length product (DLP) method for medical computed tomography (CT) dosimetry is applied on the Australian Synchrotron Imaging and Medical Beamline (IMBL). Beam quality is assessed from copper transmission measurements using image receptors, finding near 100% (20 keV), 3.3% (25 keV) and 0.5% (30–40 keV) relative contributions from third-harmonic radiation. The flat-panel-array medical image receptor is found to have a non-linear dose response curve. The amount of radiation delivered during an axial CT scan is measured as the dose in air alone, and inside cylindrical PMMA phantoms with diameters 35–160 mm for mono-energetic radiation 25–100 keV. The radiation output rate for the IMBL is comparable with that used for medical CT. Results are presented as the ratios of CT dose indices (CTDI) inside phantoms to in air with no phantom. Ratios are compared for the IMBL against medical CT where bow-tie filters shape the beam profile to reduce the absorbed dose to surface organs. CTDI ratios scale measurements in air to estimate the volumetric CTDI representing the average dose per unit length, and the dose length product representing the absorbed dose to the scanned volume. Medical CT dose calculators use the DLP, beam quality, axial collimation and helical pitch to estimate organ doses and the effective dose. The effective dose per unit DLP for medical CT is presented as a function of body region, beam energy and sample sizes from neonate to adult.Keywords
Funding Information
- University of Otago
- Australian Synchrotron
This publication has 27 references indexed in Scilit:
- X-ray phase-contrast imaging: from pre-clinical applications towards clinicsPhysics in Medicine & Biology, 2012
- RBE of low energy electrons and photonsPhysics in Medicine & Biology, 2010
- Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomographyPhysics in Medicine & Biology, 2008
- Small-Animal X-ray Dose from Micro-CTMolecular Imaging, 2004
- Quantitative determination of major systematics in synchrotron x‐ray experiments: seeing through harmonic componentsX-Ray Spectrometry, 2003
- CT-Expo - ein neuartiges Programm zur Dosisevaluierung in der CTRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2002
- Simultaneous molecular and anatomical imaging of the mousein vivoPhysics in Medicine & Biology, 2002
- Phase-contrast imaging using polychromatic hard X-raysNature, 1996
- X-ray attenuation coefficients of elements and mixturesPhysics Reports, 1981
- Statistical Error in Absorption ExperimentsPhysical Review B, 1948