Design and Characterisation of Tissue-Mimicking Gel Phantoms for Diffusion Kurtosis Imaging
Authors:
Preprint
Abstract
Purpose: The aim of this work was to create tissue-mimicking gel phantoms appropriate for diffusion kurtosis imaging (DKI) for quality assurance, protocol optimization and sequence development. Methods: A range of agar, agarose and polyvinyl alcohol phantoms with concentrations ranging from 1.0% to 3.5%, 0.5% to 3.0% and 10% to 20%, respectively, and up to 3 g of glass microspheres per 100 ml were created. Diffusion coefficients, excess kurtosis values and relaxation rates were experimentally determined. Results: The kurtosis values for the plain gels ranged from 0.05 with 95% confidence interval (CI) of $(0.029, 0.071)$ to $0.216(0.185, 0.246)$, well below the kurtosis values reported in the literature for various tissues. The addition of glass microspheres increased the kurtosis of the gels with values up to $0.523(0.465, 0.581)$ observed for gels with the highest concentration of microspheres. Repeat scans of some of the gels after more than six months of storage at room temperature indicate changes in the diffusion parameters of less than 10%. The addition of the glass microspheres reduces the apparent diffusion coefficients (ADCs) and increases the longitudinal and transverse relaxation rates but the values remain comparable to those for plain gels and tissue, with ADCs observed ranging from $818(585, 1053) \times 10^{-6}$ mm$^2$/s to $2257(2118, 2296) \times 10^{-6}$ mm$^2$/s, and R1 values ranging from $0.34(0.32, 0.35)$ 1/s to $0.51(0.50, 0.52)$ 1/s, and R2 values ranging from $9.69(9.34, 10.04)$ 1/s to $33.07(27.10, 39.04)$ 1/s. Conclusions: Glass microspheres can be used to effectively modify diffusion properties of gel phantoms and achieve a range of kurtosis values comparable to those reported for a variety of tissues.
Keywords
All Related Versions
This publication has 35 references indexed in Scilit:
- New insight into the contrast in diffusional kurtosis images: Does it depend on magnetic susceptibility?Magnetic Resonance in Medicine, 2014
- Diffusion kurtosis imaging study of prostate cancer: Preliminary findingsJournal of Magnetic Resonance Imaging, 2013
- Diffusion-weighted signal models in healthy and cancerous peripheral prostate tissues: Comparison of outcomes obtained at different b-valuesJournal of Magnetic Resonance Imaging, 2013
- A phantom for diffusion-weighted MRI (DW-MRI)Journal of Magnetic Resonance Imaging, 2013
- A simple isotropic phantom for diffusional kurtosis imagingMagnetic Resonance in Medicine, 2011
- Magnetic resonance diffusion characteristics of histologically defined prostate cancer in humansMagnetic Resonance in Medicine, 2009
- Diffusional kurtosis imaging in the lung using hyperpolarized 3HeMagnetic Resonance in Medicine, 2006
- Diffusional kurtosis imaging: The quantification of non‐gaussian water diffusion by means of magnetic resonance imagingMagnetic Resonance in Medicine, 2005
- A Phantom for diffusion‐weighted imaging of acute strokeJournal of Magnetic Resonance Imaging, 1998
- Analysis of T2 limitations and off‐resonance effects on spatial resolution and artifacts in echo‐planar imagingMagnetic Resonance in Medicine, 1990