An experimental and computational investigation of the post-yield behaviour of trabecular bone during vertebral device subsidence
Open Access
- 16 September 2012
- journal article
- research article
- Published by Springer Nature in Biomechanics and Modeling in Mechanobiology
- Vol. 12 (4) , 685-703
Abstract
Interbody fusion device subsidence has been reported clinically. An enhanced understanding of the mechanical behaviour of the surrounding bone would allow for accurate predictions of vertebral subsidence. The multiaxial inelastic behaviour of trabecular bone is investigated at a microscale and macroscale level. The post-yield behaviour of trabecular bone under hydrostatic and confined compression is investigated using microcomputed tomography-derived microstructural models, elucidating a mechanism of pressure-dependent yielding at the macroscopic level. Specifically, microstructural trabecular simulations predict a distinctive yield point in the apparent stress–strain curve under uniaxial, confined and hydrostatic compression. Such distinctive apparent stress–strain behaviour results from localised stress concentrations and material yielding in the trabecular microstructure. This phenomenon is shown to be independent of the plasticity formulation employed at a trabecular level. The distinctive response can be accurately captured by a continuum model using a crushable foam plasticity formulation in which pressure-dependent yielding occurs. Vertebral device subsidence experiments are also performed, providing measurements of the trabecular plastic zone. It is demonstrated that a pressure-dependent plasticity formulation must be used for continuum level macroscale models of trabecular bone in order to replicate the experimental observations, further supporting the microscale investigations. Using a crushable foam plasticity formulation in the simulation of vertebral subsidence, it is shown that the predicted subsidence force and plastic zone size correspond closely with the experimental measurements. In contrast, the use of von Mises, Drucker–Prager and Hill plasticity formulations for continuum trabecular bone models lead to over prediction of the subsidence force and plastic zone.Keywords
This publication has 59 references indexed in Scilit:
- Full and surface tibial cementation in total knee arthroplasty: A biomechanical investigation of stress distribution and remodeling in the tibiaClinical Biomechanics, 2012
- The Role of Fabric in the Large Strain Compressive Behavior of Human Trabecular BoneJournal of Biomechanical Engineering, 2010
- A nonlocal constitutive model for trabecular bone softening in compressionBiomechanics and Modeling in Mechanobiology, 2010
- A patient-specific finite element methodology to predict damage accumulation in vertebral bodies under axial compression, sagittal flexion and combined loadsComputer Methods in Biomechanics and Biomedical Engineering, 2008
- Prediction of strength and strain of the proximal femur by a CT-based finite element methodJournal of Biomechanics, 2007
- Cage Migration in Spondylolisthesis Treated With Posterior Lumbar Interbody Fusion Using BAK CagesSpine, 2005
- The Modified Super-Ellipsoid Yield Criterion for Human Trabecular BoneJournal of Biomechanical Engineering, 2004
- Mechanisms of uniformity of yield strains for trabecular boneJournal of Biomechanics, 2004
- Anterior lumbar fusion with paired BAK standard andpaired BAK Proximity cages: subsidence incidence, subsidence factors,and clinical outcomeThe Spine Journal, 2003
- Mechanical and Architectural Bone Adaptation in Early Stage Experimental OsteoarthritisJournal of Bone and Mineral Research, 2002