A standard SI traceable phantom suitable for qMRI: design, manufacture and characterisation
Open Access
- 14 March 2025
- journal article
- Published by IOP Publishing in Metrologia
- Vol. 62 (2) , 025008
Abstract
Quantitative MRI uses conventional clinical MRI hardware to make measurements of physical quantities. It also offers the opportunity to benchmark scanners in specific applications by characterising their measurement performance. This allows independent assessment of different scanners, products and services. Quantitation requires evaluation of the uncertainty and bias in the measurement process and as such requires a metrological framework. This process can be supported by the use of a reference object (phantom) that contains clinically relevant MRI quantities that are traceable to primary standards. Currently, there are very few traceable MRI standards available, and none cover all the forms of quantitative MRI that are being deployed clinically. The objective of this paper is to provide a metrological framework to build on and support the standardisation of qMRI through the provision of improved reference standards that are compatible with quantitative approaches. Here, we describe the design, construction, characterisation and measurement uncertainty of a traceable and metrologically quantified phantom. The phantom design is modular and comprises 30 distinct vials containing well-characterised, traceable solutions with reported uncertainty for the following MRI measurands: T 1, T 2, iron content and fat fraction.Keywords
Funding Information
- European Metrology Programme for Innovation and Research (Project 20NRM05 IMET-MRI)
This publication has 28 references indexed in Scilit:
- The perfect diagnostic imaging machine and what it means for quantitative MRI reproducibilityIPEM-Translation, 2023
- Comparative review of algorithms and methods for chemical‐shift‐encoded quantitative fat‐water imagingMagnetic Resonance in Medicine, 2023
- A standard system phantom for magnetic resonance imagingMagnetic Resonance in Medicine, 2021
- Practical guide to quantification of hepatic iron with MRIEuropean Radiology, 2019
- Recommendations towards standards for quantitative MRI (qMRI) and outstanding needsJournal of Magnetic Resonance Imaging, 2019
- Quantitative magnetic resonance imaging phantoms: A review and the need for a system phantomMagnetic Resonance in Medicine, 2017
- T2 relaxation times are increased in Skeletal muscle of DMD but not BMD patientsMuscle & Nerve, 2015
- Multicenter validation of spin‐density projection‐assisted R2‐MRI for the noninvasive measurement of liver iron concentrationMagnetic Resonance in Medicine, 2013
- Dixon techniques for water and fat imagingJournal of Magnetic Resonance Imaging, 2008