A low-cost automated flow analyzer based on low temperature co-fired ceramic and LED photometer for ascorbic acid determination
Open Access
- 21 December 2013
- journal article
- research article
- Published by Walter de Gruyter GmbH in Open Chemistry
- Vol. 12 (3) , 341-347
Abstract
An automated flow analyzer based on low temperature co-fired ceramic (LTCC), a solid-phase reactor (SPR) and a low-cost photometer was designed for ascorbic acid (AA) determination in pharmaceutical formulations. It consists of a peristaltic pump, three-way solenoid valves, SPR to chemically convert Cu(II) into Cu(I), and a LTCC device for mixing the liberated copper with bathocuproine and detection. The flow cell in the LTCC employed an ultrabright LED — photodiode photometer. The analyzer successfully determined AA in pharmaceutical formulations. The analytical curve from 8.5×10−6 to 7.0×10−4 M gave a detection limit of 7.0×10−7 M and a RSD of 2.1% for a 2.0×10−4 M AA solution (n = 10). A high sampling frequency of 102 h−1 and low reagent and sample consumption (150 µL) resulted.Keywords
This publication has 36 references indexed in Scilit:
- Automated Determination of Captopril by Flow and Sequential Injection Analysis: A ReviewAnalytical Letters, 2011
- Use of Barium Chloranilate Solid-Phase Reactor for the Determination of Sulfate in Natural Water Samples Exploiting Long Pathlength Spectrophotometry and MulticommutationAnalytical Letters, 2011
- A compact miniaturized continuous flow system for the determination of urea content in milkAnalytical and Bioanalytical Chemistry, 2010
- Multi-commutation in spectrometryTrAC Trends in Analytical Chemistry, 2009
- Automated sample preparation coupled to sequential injection chromatography: On-line filtration and dilution protocols prior to separationJournal of Pharmaceutical and Biomedical Analysis, 2009
- Amperometric determination of cyanides at the low ppb level by automated preconcentration based on gas diffusion coupled to sequential injection analysisTalanta, 2009
- Flow-injection analysis with a small cylindrical wire type electrodeOpen Chemistry, 2004
- Multicommutation in flow analysis: concepts, applications and trendsAnalytica Chimica Acta, 2002
- Automated Monosegmented Flow Analyser. Determination of Glucose, Creatinine and UreaThe Analyst, 1997
- Multicommutation in flow analysis. Part 1. Binary sampling: concepts, instrumentation and spectrophotometric determination of iron in plant digestsAnalytica Chimica Acta, 1994