Comparison of the force applied on oral structures during intubation attempts by novice physicians between the Macintosh direct laryngoscope, Airway Scope and C-MAC PM: a high-fidelity simulator-based study
Open Access
Abstract
Objective We examined whether the use of Airway Scope (AWS) and C-MAC PM (C-MAC) decreased the force applied on oral structures during intubation attempts as compared with the force applied with the use of Macintosh direct laryngoscope (DL). Design Prospective cross-over study. Participants A total of 35 novice physicians participated. Interventions We used 6 simulation scenarios based on the difficulty of intubation and intubation devices. Outcome measures Our primary outcome measures were the maximum force applied on the maxillary incisors and tongue during intubation attempts, measured by a high-fidelity simulator. Results The maximum force applied on maxillary incisors was higher with the use of the C-MAC than with the DL and AWS in the normal airway scenario (DL, 26 Newton (N); AWS, 18 N; C-MAC, 52 N; pConclusions The use of C-MAC, compared with the DL and AWS, was associated with the higher maximum force applied on maxillary incisors during intubation attempts. In contrast, the use of video laryngoscopes was associated with the lower force applied on the tongue in both airway scenarios, compared with the DL. Our study was a simulation-based study, and further research on living patients would be warranted.Keywords
This publication has 24 references indexed in Scilit:
- Video laryngoscopy improves intubation success and reduces esophageal intubations compared to direct laryngoscopy in the medical intensive care unitCritical Care, 2013
- The Pentax airway scope versus the Macintosh laryngoscope: Comparison of hemodynamic responses and concentrations of plasma norepinephrine to tracheal intubationKorean Journal of Anesthesiology, 2013
- Investigation of the freely available easy-to-use software ‘EZR’ for medical statisticsBone Marrow Transplantation, 2012
- Measurement of forces applied during Macintosh direct laryngoscopy compared with GlideScope ® videolaryngoscopy*Anaesthesia, 2012
- Force and pressure distribution using Macintosh and GlideScope laryngoscopes in normal and difficult airways: a manikin studyBritish Journal of Anaesthesia, 2011
- Forces applied by the laryngoscope blade onto the tongue during intubation attempts: a comparison between Mackintosh, AirTraq and Pentax AWS in a mannequin studyEuropean Journal of Anaesthesiology, 2011
- Traumatic dental injuries during anaesthesia: part I: clinical evaluationDental Traumatology, 2010
- A comparison of McGrath and Macintosh laryngoscopes in novice users: a manikin studyAnaesthesia, 2009
- A comparison of the Glidescope®, Pentax AWS®, and Macintosh laryngoscopes when used by novice personnel: a manikin studyCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 2009
- Forces Applied to the Maxillary Incisors During Video-Assisted IntubationAnesthesia & Analgesia, 2009