A Multi-Center Trained Residual Neural Network for Robust Classification of Atrial High-Rate Episodes in Remotely Monitored Pacemakers and Defibrillators
Authors:
- Lars van KrimpenLKLars van Krimpen
- Arlene JohnAJArlene John
- Anand ThiyagarajahATAnand Thiyagarajah
- Tanner CarbonatiTCTanner Carbonati
- Benjamin SacristanBSBenjamin Sacristan
- Karim BenaliKBKarim Benali
- Antoine Da CostaACAntoine Da Costa
- Pierre MondolyPMPierre Mondoly
- Rémi ChauvelRCRémi Chauvel
- Romain EschalierRERomain Eschalier
- Josselin DuchateauJDJosselin Duchateau
- Remi DuboisRDRemi Dubois
Open Access
Abstract
Remote monitoring of pacemakers and defibrillators increases patient safety but also increases clinical workload. Review of atrial high-rate episodes is particularly demanding as episodes can contain atrial tachycardia or atrial fibrillation (AT/AF), noise, or far-field oversensing (FFO). Automatic review of atrial high-rate episodes by an Artificial Intelligence (AI) model can decrease the workload of remote monitoring, provided it maintains high sensitivity for true atrial tachycardia. A residual network is trained using a center-level fourfold cross validation. The four resulting models achieved a precision of 97.2–99.4% for AT/AF, 93.1–97.7% for noise, and 75.4–94.4% for FFO, while maintaining high sensitivity 98.9–99.3% for AT/AF. The four models were combined through averaging prediction probabilities to create an ensemble model. Thresholding ensemble predictions with probability > 95% resulted in a robust ensemble model that made only two errors (<0.1%) after reviewing 3925 episodes (91.9%) of the total 4271 episodes. This shows how AI models can reliably assist in remote monitoring. Future research should be aimed at classification models for other episode types and clinical validation of AI models to assist remote monitoring of pacemakers and defibrillators.Keywords
This publication has 20 references indexed in Scilit:
- Atrial fibrillation detection performance of an insertable cardiac monitor: Results from an Assert-IQ post-market clinical study and a novel artificial intelligence algorithmHeart Rhythm O2, 2025
- Arrhythmia care in ESC member countries: the 2025 ESC-EHRA atlas on heart rhythm disordersEP Europace, 2025
- Impact of Artificial Intelligence-Enhanced Insertable Cardiac Monitors on Device Clinic Workflow and Resource UtilizationJACC: Advances, 2025
- PO-02-175 AI ENABLES SIGNIFICANT REDUCTION OF CLINIC REVIEW BURDEN FOR LEGACY ICMSHeart Rhythm, 2024
- Apixaban for Stroke Prevention in Subclinical Atrial FibrillationNew England Journal of Medicine, 2024
- Anticoagulation with Edoxaban in Patients with Atrial High-Rate EpisodesNew England Journal of Medicine, 2023
- Artificial intelligence for detection of ventricular oversensing: Machine learning approaches for noise detection within nonsustained ventricular tachycardia episodes remotely transmitted by pacemakers and implantable cardioverter-defibrillatorsHeart Rhythm, 2023
- 2023 HRS/EHRA/APHRS/LAHRS Expert Consensus Statement on Practical Management of the Remote Device ClinicEP Europace, 2023
- Remote monitoring of pacemakersArchives of Cardiovascular Diseases, 2021
- The CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) TrialJACC, 2011