Mode of Breathing—Tidal or Slow and Deep—through the I-neb Adaptive Aerosol Delivery (AAD) System Affects Lung Deposition of 99m Tc-DTPA
- 1 April 2010
- journal article
- research article
- Published by SAGE Publications in Journal of Aerosol Medicine and Pulmonary Drug Delivery
- Vol. 23 (s1) , S37-37
Abstract
Background: The I-neb AAD System was designed to deliver aerosol with two different breathing pattern algorithms: the Tidal Breathing Mode (TBM) and the Target Inhalation Mode (TIM). For the purpose of the study, the TBM breathing pattern algorithm was set to guide the subjects to inhalation during tidal breathing with aerosol pulsed during 50–80% of the time spent on inhalation, whereas the TIM breathing pattern was set to guide the subject to a slow and deep inhalation of up to ∼9 sec with aerosol pulsed for up to 7 sec, leaving 2 sec for particle deposition in the lungs. In TIM, the inspiratory flow was guided to ∼20 L/min through a built-in resistance in the mouthpiece. Methods: We have, in a randomized, open-label, crossover study of 12 healthy subjects evaluated lung deposition following administration of a radiolabeled aerosol from the I-neb AAD System with the TBM and TIM breathing patterns. Results: The results showed that mean lung deposition was significantly higher when using the I-neb AAD System with the TIM breathing pattern (73.3%) than with the TBM breathing pattern (62.8%). The mean exhaled fractions were low (Conclusions: The results of the present study showed that lung deposition with the slow and deep inhalation achieved through the I-neb AAD System in TIM was superior to the lung deposition achieved during tidal breathing in TBM. With the combination of high lung deposition, almost no loss of aerosol during exhalation, and short nebulization time the I-neb AAD System with the TIM breathing pattern should be of special value to patients who require multiple daily dosing of aerosolized medication, are using drugs that should not be wasted into the room air, or would benefit from a more efficient delivery system.Keywords
This publication has 11 references indexed in Scilit:
- Evaluation of the Target Inhalation Mode (TIM) Breathing Maneuver in Simulated Nebulizer Therapy in Patients with Cystic FibrosisJournal of Aerosol Medicine and Pulmonary Drug Delivery, 2010
- Lung deposition of inhaled α1-proteinase inhibitor in cystic fibrosis and α1-antitrypsin deficiencyEuropean Respiratory Journal, 2009
- Adaptive Aerosol Delivery (AAD®) Technology2008
- Delivery of Iloprost Inhalation Solution With the HaloLite, Prodose, and I-neb Adaptive Aerosol Delivery Systems: An In Vitro StudyRespiratory Care, 2007
- History of Aerosol Therapy: Liquid Nebulization to MDIs to DPIsRespiratory Care, 2005
- Adaptive Aerosol Delivery (AAD®) technologyExpert Opinion on Drug Delivery, 2004
- Systemic availability and lung deposition of budesonide via three different nebulizers in adultsAnnals of Allergy, Asthma & Immunology, 2003
- HUMAN TRACHEOBRONCHIAL DEPOSITION AND EFFECT OF A HISTAMINE AEROSOL INHALED BY EXTREMELY SLOW INHALATIONSJournal of Aerosol Science, 1999
- HUMAN TRACHEOBRONCHIAL DEPOSITION AND EFFECT OF A CHOLINERGIC AEROSOL INHALED BY EXTREMELY SLOW INHALATIONSExperimental Lung Research, 1999
- Effects of various inhalation modes on the deposition of radioactive pressurized aerosols.1982