Assessment of Heating on Titanium Alloy Cerebral Aneurysm Clips during 7T MRI
- 23 June 2022
- journal article
- research article
- Published by American Society of Neuroradiology (ASNR) in American Journal of Neuroradiology
- Vol. 43 (7) , 972-977
Abstract
BACKGROUND AND PURPOSE: Patients with cerebral aneurysms often undergo MR imaging after microsurgical clipping. Ultra-high-field MR imaging at 7T may provide high diagnostic capability in such clinical situations. However, titanium alloy clips have safety issues such as adverse interactions with static magnetic fields and radiofrequency-induced heating during 7T MR imaging. The purpose of this study was to quantitatively assess temperature increases on various types of titanium alloy aneurysm clips during 7T MR imaging. MATERIALS AND METHODS: Five types of titanium alloy aneurysm clips were tested, including combinations of short, long, straight, angled, and fenestrated types. Each clip was set in a phantom filled with gelled saline mixed with polyacrylic acid and underwent 7T MR imaging with 3D T1WI with a spoiled gradient recalled acquisition in the steady-state technique. Temperature was chronologically measured at the tips of the clip blade and head, angled part of the clip, and 5 mm from the tip of the clip head using MR imaging–compatible fiber-optic thermometers. RESULTS: Temperature increases at all locations for right-angled and short straight clips were 1°C. Temperature increases at all locations for the long straight clip were >2°C. CONCLUSIONS: Temperature increases on the right-angled and short straight clips remained below the regulatory limit during 7T MR imaging, but temperature increases on the 45° angled, straight fenestrated, and long straight clips exceeded this limit.Keywords
This publication has 25 references indexed in Scilit:
- Assessment of MRI Issues at 7 T for 28 Implants and Other ObjectsAmerican Journal of Roentgenology, 2014
- Perforating arteries originating from the posterior communicating artery: a 7.0-Tesla MRI studyEuropean Radiology, 2009
- First Clinical Study on Ultra-High-Field MR Imaging in Patients with Multiple Sclerosis: Comparison of 1.5T and 7TAmerican Journal of Neuroradiology, 2009
- Intracranial time‐of‐flight MR angiography at 7T with comparison to 3TJournal of Magnetic Resonance Imaging, 2007
- High-Field-Strength Magnetic ResonanceTopics in Magnetic Resonance Imaging, 2007
- Actual flip‐angle imaging in the pulsed steady state: A method for rapid three‐dimensional mapping of the transmitted radiofrequency fieldMagnetic Resonance in Medicine, 2006
- Evaluation of specific absorption rate as a dosimeter of MRI‐related implant heatingJournal of Magnetic Resonance Imaging, 2004
- Magnetic resonance angiography with ultrashort echo times reduces the artefact of aneurysm clipsNeuroradiology, 2002
- Symptomatic vasospasm after subarachnoid haemorrhage: assessment of brain damage by diffusion and perfusion-weighted MRI and single-photon emission computed tomographyNeuroradiology, 2002
- Signal‐to‐noise ratio and absorbed power as functions of main magnetic field strength, and definition of “90°” RF pulse for the head in the birdcage coilMagnetic Resonance in Medicine, 2001