Abstract
A 54-year-old man with Parkinson’s disease presented with gait instability and difficulty charging his rechargeable deep brain stimulation (DBS) pulse generator. Chest radiography revealed horizontal rotation and anterior-to-posterior inversion of the device within the subcutaneous pocket, which was not seen in prior imaging. Surgical revision confirmed the device malorientation, and repositioning restored charging efficiency. This case highlights that pulse generator rotation, a rare but relevant hardware complication, can cause charging difficulties despite normal device interrogation. It underscores the importance of considering mechanical causes and using chest imaging in postoperative assessments.Keywords
Funding Information
- Aerospace Medical Health Technology Group Co., LTD. (2025YK24)
- National Natural Science Foundation of China (62171003)
This publication has 5 references indexed in Scilit:
- Recharging Difficulty With Pulse Generator After Deep Brain Stimulation: A Case Series of Five PatientsFrontiers in Neuroscience, 2021
- A systematic review of Twiddler’s syndrome: a hardware-related complication of deep brain stimulationNeurosurgical Review, 2021
- Rechargeable Pacemaker Technology in Deep Brain Stimulation: A Step Forward, But Not for EveryoneMovement Disorders Clinical Practice, 2021
- Surgical and Hardware Complications of Deep Brain Stimulation—A Single Surgeon Experience of 519 Cases Over 20 YearsNeuromodulation: Technology at the Neural Interface, 2021
- Fixed-Life or Rechargeable Battery for Deep Brain Stimulation: Which Do Patients Prefer?Neuromodulation: Technology at the Neural Interface, 2018