A 3D stack‐of‐spirals approach for rapid hyperpolarized 129Xe ventilation mapping in pediatric cystic fibrosis lung disease
- 25 November 2022
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 89 (3) , 1083-1091
Abstract
Purpose: To demonstrate the feasibility of a rapid 3D stack‐of‐spirals (3D‐SoS) imaging acquisition for hyperpolarized 129Xe ventilation mapping in healthy pediatric participants and pediatric cystic fibrosis (CF) participants, in comparison to conventional Cartesian multislice (2D) gradient‐recalled echo (GRE) imaging.Methods: The 2D‐GRE and 3D‐SoS acquisitions were performed in 13 pediatric participants (5 healthy, 8 CF) during separate breath‐holds. Images from both sequences were compared on the basis of ventilation defect percent (VDP) and other measures of image similarity. The nadir of transient oxygen saturation (SpO2) decline due to xenon breath‐holding was measured with pulse oximetry, and expressed as a percent change relative to baseline.Results: 129Xe ventilation images were acquired in a breath‐hold of 1.2–1.8 s with the 3D‐SoS sequence, compared to 6.2–8.8 s for 2D‐GRE. Mean ± SD VDP measures for 2D‐GRE and 3D‐SoS sequences were 5.02 ± 1.06% and 5.28 ± 1.08% in healthy participants, and 18.05 ± 8.26% and 18.75 ± 6.74% in CF participants, respectively. Across all participants, the intraclass correlation coefficient of VDP measures for both sequences was 0.98 (95% confidence interval: 0.94–0.99). The percent change in SpO2 was reduced to −2.1 ± 2.7% from −5.2 ± 3.5% with the shorter 3D‐SoS breath‐hold.Conclusion: Hyperpolarized 129Xe ventilation imaging with 3D‐SoS yielded images approximately five times faster than conventional 2D‐GRE, reducing SpO2 desaturation and improving tolerability of the xenon administration. Analysis of VDP and other measures of image similarity demonstrate excellent agreement between images obtained with both sequences. 3D‐SoS holds significant potential for reducing the acquisition time of hyperpolarized 129Xe MRI, and/or increasing spatial resolution while adhering to clinical breath‐hold constraints.Keywords
Funding Information
- Canadian Institutes of Health Research (PJT‐1530399)
- Cystic Fibrosis Foundation
This publication has 27 references indexed in Scilit:
- ESPIRiT—an eigenvalue approach to autocalibrating parallel MRI: Where SENSE meets GRAPPAMagnetic Resonance in Medicine, 2013
- Hyperpolarized 129Xe MRI of the human lungJournal of Magnetic Resonance Imaging, 2013
- SPIRiT: Iterative self‐consistent parallel imaging reconstruction from arbitrary k‐spaceMagnetic Resonance in Medicine, 2010
- Sparse MRI: The application of compressed sensing for rapid MR imagingMagnetic Resonance in Medicine, 2007
- Functional lung imaging using hyperpolarized gas MRIJournal of Magnetic Resonance Imaging, 2007
- Comparison between 2D and 3D gradient‐echo sequences for MRI of human lung ventilation with hyperpolarized 3HeMagnetic Resonance in Medicine, 2004
- Image quality assessment: from error visibility to structural similarityIEEE Transactions on Image Processing, 2004
- Fast 3D imaging using variable‐density spiral trajectories with applications to limb perfusionMagnetic Resonance in Medicine, 2003
- Nonuniform fast fourier transforms using min-max interpolationIEEE Transactions on Signal Processing, 2003
- Measures of the Amount of Ecologic Association Between SpeciesEcology, 1945