Abstract
Time-of-flight positron emission tomography (TOF-PET) scanners demand electronics that are power-efficient, low-noise, cost-effective, and possess a large bandwidth. Recent developments have demonstrated sub-100 ps time resolution with elevated power consumption per channel, rendering this unfeasible to build a scanner. In this work, we evaluate the performance for TOF-PET of the FastIC front-end using different scintillators and silicon photomultipliers (SiPMs). FastIC is an 8-channel ASIC developed in CMOS 65 nm capable of measuring the energy and the arrival time of a detected pulse with 12 mW per channel. Using Hamamatsu SiPMs (S13360-3050PE) coupled to LSO:Ce:0.2%Ca crystals of 2 × 2 × 3 mm and LYSO:Ce:0.2%Ca of 3.13 × 3.13 × 20 mm, we measured a CTR of (95 ± 3) ps and 156 ± 4) ps FWHM, respectively. With FBK NUV-HD LF2 M0 SiPMs coupled to the same crystals, we obtained a CTR of (76 ± 2) ps and (127 ± 3) ps FWHM. We employed FastIC with a TlCl pure Cherenkov emitter, demonstrating time resolutions comparable to those achieved with high-power-consuming electronics. These findings shows that FastIC represents a cost-effective alternative that can significantly enhance the time resolution of current TOF-PET systems while maintaining low power consumption.

This publication has 54 references indexed in Scilit: