Toward Sub-100 ps TOF-PET Systems Employing the FastIC ASIC With Analog SiPMs
Open Access
- 14 June 2024
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Radiation and Plasma Medical Sciences
- Vol. 8 (7) , 718-733
Abstract
Time-of-flight positron emission tomography (TOF-PET) scanners demand electronics that are power-efficient, low-noise, cost-effective, and possess a large bandwidth. Recent developments have demonstrated sub-100 ps time resolution with elevated power consumption per channel, rendering this unfeasible to build a scanner. In this work, we evaluate the performance for TOF-PET of the FastIC front-end using different scintillators and silicon photomultipliers (SiPMs). FastIC is an 8-channel ASIC developed in CMOS 65 nm capable of measuring the energy and the arrival time of a detected pulse with 12 mW per channel. Using Hamamatsu SiPMs (S13360-3050PE) coupled to LSO:Ce:0.2%Ca crystals of 2 × 2 × 3 mm and LYSO:Ce:0.2%Ca of 3.13 × 3.13 × 20 mm, we measured a CTR of (95 ± 3) ps and 156 ± 4) ps FWHM, respectively. With FBK NUV-HD LF2 M0 SiPMs coupled to the same crystals, we obtained a CTR of (76 ± 2) ps and (127 ± 3) ps FWHM. We employed FastIC with a TlCl pure Cherenkov emitter, demonstrating time resolutions comparable to those achieved with high-power-consuming electronics. These findings shows that FastIC represents a cost-effective alternative that can significantly enhance the time resolution of current TOF-PET systems while maintaining low power consumption.Keywords
This publication has 54 references indexed in Scilit:
- Hybrid total-body pet scanners—current status and future perspectivesEuropean Journal of Nuclear Medicine and Molecular Imaging, 2021
- Overview on the main parameters and technology of modern Silicon PhotomultipliersNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2019
- The new opportunities for high time resolution clinical TOF PETClinical and Translational Imaging, 2019
- Approximations of coincidence time resolution models of scintillator detectors with leading edge discriminatorNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2018
- BGO as a hybrid scintillator / Cherenkov radiator for cost-effective time-of-flight PETPhysics in Medicine & Biology, 2017
- Bismuth germanate coupled to near ultraviolet silicon photomultipliers for time-of-flight PETPhysics in Medicine & Biology, 2016
- Fundamental limits of scintillation detector timing precisionPhysics in Medicine & Biology, 2014
- Time resolution deterioration with increasing crystal length in a TOF-PET systemNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2014
- The lower bound on the timing resolution of scintillation detectorsPhysics in Medicine & Biology, 2012
- NINO: an ultra-fast and low-power front-end amplifier/discriminator ASIC designed for the multigap resistive plate chamberNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2004