Adaptive spatial calibration of a 3D ultrasound system
Authors:
- Alex HartovAHAlex Hartov
- Keith PaulsenKPKeith Paulsen
- Songbai JiSJSongbai Ji
- Kathryn FontaineKFKathryn Fontaine
- Marie‐Laure FuronMFMarie‐Laure Furon
- Andrea BorsicABAndrea Borsic
- David RobertsDRDavid Roberts
- 22 April 2010
- journal article
- Published by Wiley in Medical Physics
- Vol. 37 (5) , 2121-2130
Abstract
Purpose: The authors present a method devised to calibrate the spatial relationship between a 3D ultrasound scanhead and its tracker completely automatically and reliably. The user interaction is limited to collecting ultrasound data on which the calibration is based.Methods: The method of calibration is based on images of a fixed plane of unknown location with respect to the 3D tracking system. This approach has, for advantage, to eliminate the measurement of the plane location as a source of error. The devised method is sufficiently general and adaptable to calibrate scanheads for 2D images and 3D volume sets using the same approach. The basic algorithm for both types of scanheads is the same and can be run unattended fully automatically once the data are collected. The approach was devised by seeking the simplest and most robust solutions for each of the steps required. These are the identification of the plane intersection within the images or volumes and the optimization method used to compute a calibration transformation matrix. The authors use adaptive algorithms in these two steps to eliminate data that would otherwise prevent the convergence of the procedure, which contributes to the robustness of the method.Results: The authors have run tests amounting to 57 runs of the calibration on two a scanhead that produce 3D imaging volumes, at all the available scales. The authors evaluated the system on two criteria: Robustness and accuracy. The program converged to useful values unattended for every one of the tests (100%). Its accuracy, based on the measured location of a reference plane, was estimated to be for all tests combined.Conclusions: The system presented is robust and allows unattended computations of the calibration parameters required for freehand tracked ultrasound based on either 2D or 3D imaging systems.Keywords
This publication has 30 references indexed in Scilit:
- Initial Experience with an Ultrasound-Integrated Single-Rack Neuronavigation SystemActa Neurochirurgica, 2001
- Comparison of intraoperative MR imaging and 3D-navigated ultrasonography in the detection and resection control of lesionsNeurosurgical Focus, 2001
- Quantification of, Visualization of, and Compensation for Brain Shift Using Intraoperative Magnetic Resonance ImagingNeurosurgery, 2000
- Intraoperative ultrasound for guidance and tissue shift correction in image‐guided neurosurgeryMedical Physics, 2000
- Ultrasound/MRI Overlay with Image Warping for Neurosurgery2000
- Model-updated image guidance: initial clinical experiences with gravity-induced brain deformationIEEE Transactions on Medical Imaging, 1999
- Intraoperative Brain Shift and Deformation: A Quantitative Analysis of Cortical Displacement in 28 CasesNeurosurgery, 1998
- Intraoperative US in interactive image-guided neurosurgery.RadioGraphics, 1998
- Intraoperative Three-Dimensional Ultrasonography: An Approach to Register Brain Shift using Multidimensional Image Processingmin - Minimally Invasive Neurosurgery, 1998
- Frameless stereotactic ultrasonography: Method and applicationsComputerized Medical Imaging and Graphics, 1994