Design and performance evaluation of a slit‐slat camera for 2D prompt gamma imaging in proton therapy monitoring: A Monte Carlo simulation study
Authors:
- 15 March 2023
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 50 (6) , 3701-3718
Abstract
PurposeWe investigated the design of a prompt gamma camera for real-time dose delivery verification and the partial mitigation of range uncertainties. MethodsA slit slat (SS) camera was optimized using the trade-off between the signal-to-noise ratio and spatial resolution. Then, using the GATE Monte Carlo package, the camera performances were estimated by means of target shifts, beam position quantification, changing the camera distance from the beam, and air cavity inserting. A homogeneous PMMA phantom and the air gaps induced PMMA phantom were used. The air gaps ranged from 5 mm to 30 mm by 5 mm increments were positioned in the middle of the beam range. To reduce the simulation time, phase space scoring was used. The batch method with five realizations was used for stochastic error calculations. ResultsThe system's detection efficiency was 1.1x10-4PGsEmittedPGs(1.8x10-5 PGs/proton) for a 10 x 20 cm(2) detector (source-to-collimator distance = 15.0 cm). Axial and transaxial resolutions were 23 mm and 18 mm, respectively. The SS camera estimated the range as 69.0 +/- 3.4 (relative stochastic error 1-sigma is 5%) and 67.6 +/- 1.8 mm (2.6%) for the real range of 67.0 mm for 10(7) and 10(8) protons of 100 MeV, respectively. Considering 160 MeV, these values are 155.5 +/- 3.1 (2%) and 152.2 +/- 2.0 mm (1.3%) for the real range of 152.0 mm for 10(7) and 10(8) protons, respectively. Considering phantom shift, for a 100 MeV beam, the precision of the quantification (1-sigma) in the axial and lateral phantom shift estimation is 2.6 mm and 1 mm, respectively. Accordingly, the axial and lateral quantification precisions were 1.3 mm and 1 mm for a 160 MeV beam, respectively. Furthermore, the quantification of an air gap formulated as gapdet=0.98xgapreal and gap(real) are the estimated and real air gap, respectively. The precision of the air gap quantification is 1.6 mm (1 sigma). Moreover, 2D PG images show the trajectory of the proton beam through the phantom. ConclusionThe proposed slit-slat imaging systems can potentially provide a real-time, in-vivo, and non-invasive treatment monitoring method for proton therapy.Keywords
This publication has 53 references indexed in Scilit:
- In vivoproton range verification: a reviewPhysics in Medicine & Biology, 2013
- Prompt gamma imaging with a slit camera for real-time range control in proton therapyPhysics in Medicine & Biology, 2012
- Range uncertainties in proton therapy and the role of Monte Carlo simulationsPhysics in Medicine & Biology, 2012
- Real-time prompt gamma monitoring in spot-scanning proton therapy using imaging through a knife-edge-shaped slitPhysics in Medicine & Biology, 2011
- Monte Carlo patient study on the comparison of prompt gamma and PET imaging for range verification in proton therapyPhysics in Medicine & Biology, 2011
- GATE V6: a major enhancement of the GATE simulation platform enabling modelling of CT and radiotherapyPhysics in Medicine & Biology, 2011
- Optimizing a three-stage Compton camera for measuring prompt gamma rays emitted during proton radiotherapyPhysics in Medicine & Biology, 2010
- ROOT — A C++ framework for petabyte data storage, statistical analysis and visualizationComputer Physics Communications, 2009
- Pinhole Camera Measurements of Prompt Gamma-rays for Detection of Beam Range Variation in Proton TherapyJournal of the Korean Physical Society, 2009
- Patient Study of In Vivo Verification of Beam Delivery and Range, Using Positron Emission Tomography and Computed Tomography Imaging After Proton TherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2007