Differential sampling of AC waveforms based on a programmable Josephson voltage standard using a high-precision sampler
- 7 January 2022
- journal article
- research article
- Published by IOP Publishing in Metrologia
- Vol. 59 (1) , 015006
Abstract
A high-precision sampler, Fluke 8588A multimeter in the sampling mode, was utilised to perform differential sampling of AC waveforms with a programmable Josephson voltage standard. The systematic error on the differential sampling, induced by the inherent voltage-response characteristics and built-in low-pass filter of the sampler, was estimated. Experimental results and numerical simulations revealed that the sampler could be used for reliable differential sampling of AC waveforms at frequencies up to several kilohertz, with an appropriate number of the voltage steps per the waveform period, when the input bandwidth was set to 3 MHz. In addition, the sampler was compared to an integrating sampler, Keysight 3458A, now widely used for differential sampling. At 62.5 Hz, a key frequency in the future on-site key comparison of the differential sampling on AC voltage, the difference in root mean square (RMS) amplitudes obtained by the differential sampling using the two different samplers is approximately 150 nV V−1 due to the systematic error caused by the limited bandwidth of 150 kHz for the integrating sampler.Keywords
Funding Information
- Korea Research Institute of Standards and Science (KRISS-2021-GP2021-0001)
This publication has 22 references indexed in Scilit:
- Development of a PJVS based primary AC power standard2016
- Uncertainty Evaluation of a 10 V RMS Sampling Measurement System Using the AC Programmable Josephson Voltage StandardIEEE Transactions on Instrumentation and Measurement, 2015
- Junction Yield Analysis for 10 V Programmable Josephson Voltage Standard DevicesIEEE Transactions on Applied Superconductivity, 2014
- An ac quantum voltmeter based on a 10 V programmable Josephson arrayMetrologia, 2013
- NIST 10 V Programmable Josephson Voltage Standard SystemIEEE Transactions on Instrumentation and Measurement, 2011
- Analog-to-digital conversion for low-frequency waveforms based on the Josephson voltage standardMeasurement Science and Technology, 2010
- Precision sampling measurements using ac programmable Josephson voltage standardsReview of Scientific Instruments, 2008
- Classical Nonquantum AC Power Measurements With Uncertainties Approaching 1 ${\rm \mu}\hbox{W/VA}$IEEE Transactions on Instrumentation and Measurement, 2007
- Direct Comparison of Josephson Waveforms Using an AC Quantum VoltmeterIEEE Transactions on Instrumentation and Measurement, 2007
- Evaluation of the synchronous generation and sampling techniqueIEEE Transactions on Instrumentation and Measurement, 2003