Electroanalytical Device for Cadmium Speciation in Waters. Part 2: Automated System Development and Cadmium Semicontinuous Monitoring
- 21 May 2010
- journal article
- research article
- Published by Wiley in Electroanalysis
- Vol. 22 (11) , 1159-1165
Abstract
A full‐automated electroanalytical device is presented here. It is composed of an electrochemically adapted automation module and a competitive screen‐printed sensor modified ex situ by a very low quantity of mercury. The contribution of this automated apparatus to analyses is described in detail. Coupled with an UV‐photolysis unit, it has permitted the rapid detection of both total and electrochemically accessible cadmium in solutions containing 10 mg L−1 of humic substances without addition of reagents. Semicontinuous monitoring of tap water doped with 5.3 μg L−1 of cadmium, in presence of traces of lead and copper, has been successfully achieved for 7 days.Keywords
This publication has 35 references indexed in Scilit:
- Electroanalytical Device for Cadmium Speciation in Waters. Part 1: Development and Characterization of a Reliable Screen‐Printed SensorElectroanalysis, 2010
- A review on sequential injection methods for water analysisAnalytica Chimica Acta, 2009
- Electrochemical SensorsAnalytical Chemistry, 2008
- Square wave adsorptive cathodic stripping voltammetry automated by sequential injection analysis: Potentialities and limitations exemplified by the determination of methyl parathion in water samplesAnalytica Chimica Acta, 2008
- Determination of labile trace metals with screen-printed electrode modified by a crown-ether based membraneAnalytica Chimica Acta, 2006
- Miniaturised stripping-based carbon modified sensor for in field analysis of heavy metalsAnalytica Chimica Acta, 2005
- An automated monitor to determine trace metal particle/dissolved interactions in natural watersAnalytica Chimica Acta, 2004
- Possibilities for the automated determination of trace concentrations of uranium in water samples by adsorptive stripping voltammetryAnalytica Chimica Acta, 1997
- Automated voltammetric system for shipboard determination of metal speciation in sea waterAnalytica Chimica Acta, 1994
- Flow injection system for stripping voltammetryAnalytica Chimica Acta, 1985