Suitability of Polyacrylamide-Based Dosimetric Gel for Proton and Carbon Ion Beam Geometric Characterization
Authors:
- Riccardo BrambillaRBRiccardo Brambilla
- Luca TrombettaLTLuca Trombetta
- Gabriele MaguglianiGMGabriele Magugliani
- Stefania RussoSRStefania Russo
- Alessia BazaniABAlessia Bazani
- Eleonora RossiEREleonora Rossi
- Eros MossiniEMEros Mossini
- Elena MacerataEMElena Macerata
- Francesco GalluccioFGFrancesco Galluccio
- Mario MarianiMMMario Mariani
- Mario CioccaMCMario Ciocca
Open Access
Abstract
Experimental measurement of dose distributions is a pivotal step in the quality assurance of radiotherapy treatments, especially for those relying on high delivery accuracy such as hadron therapy. This study investigated the response of a polymer gel dosimeter to determine its suitability in performing geometric beam characterizations for hadron therapy under high-quenching conditions. Different extraction energies of proton and carbon ion beams were considered. Gel dose–response linearity and long-term stability were confirmed through optical measurements. Gel phantoms were irradiated with pencil beams and analyzed via magnetic resonance imaging. A multi-echo T2-weighted sequence was used to reconstruct depth–dose profiles and transversal distributions acquired by the gels, which were benchmarked against reference data. As expected, a response-quenching effect in the Bragg peak region was noted. Nonetheless, the studied gel formulation proved reliable in acquiring the geometric characteristics of the beams, even without correcting for the quenching effect. Indeed, depth–dose distributions acquired by the gels showed an excellent agreement with measured particle range with respect to reference values, with mean discrepancies of 0.5 ± 0.2 mm. Single-spot transverse FWHM values at increasing depths also presented an average agreement within 1 mm with values determined with radiochromic films, thus supporting the excellent spatial resolving capabilities of the dosimetric gel.Keywords
This publication has 40 references indexed in Scilit:
- Pilot scale validation campaign of gel dosimetry for pre-treatment quality assurance in stereotactic radiotherapyPhysica Medica, 2023
- Empirical quenching correction in radiochromic silicone-based three-dimensional dosimetry of spot-scanning proton therapyPhysics and Imaging in Radiation Oncology, 2021
- Gel dosimetry for three dimensional proton range measurements in anthropomorphic geometriesZeitschrift für Medizinische Physik, 2019
- A systematic review of clinical applications of polymer gel dosimeters in radiotherapyApplied Radiation and Isotopes, 2019
- Experimental methods in chemical engineering: Ultraviolet visible spectroscopy—UV‐VisThe Canadian Journal of Chemical Engineering, 2018
- Characteristics of a novel polymer gel dosimeter formula for MRI scanning: Dosimetry, toxicity and temporal stability of responsePhysica Medica, 2016
- Radiochromic film dosimetry: Past, present, and futurePhysica Medica, 2011
- Radiological properties of the PRESAGE and PAGAT polymer dosimetersApplied Radiation and Isotopes, 2008
- Optical analysis of gel dosimeters: Comparison of Fricke and normoxic polymer gelsNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 2007
- Mathematical analysis and experimental investigation of noise in quantitative magnetic resonance imaging applied in polymer gel dosimetrySignal Processing, 1998