Risk Factors Associated with Hemodialysis Central Venous Catheter Malfunction; A Retrospective Analysis of a Randomized Controlled Trial
Authors:
- David R WardDWDavid R Ward
- Louise M MoistLMLouise M Moist
- Jennifer M MacRaeJMJennifer M MacRae
- Nairne Scott-DouglasNSNairne Scott-Douglas
- Jianguo ZhangJZJianguo Zhang
- Marcello TonelliMTMarcello Tonelli
- Charmaine E LokCLCharmaine E Lok
- Steven D SorokaSSSteven D Soroka
- Brenda R HemmelgarnBHBrenda R Hemmelgarn
Open Access
- 1 January 2014
- journal article
- Published by SAGE Publications in Canadian Journal of Kidney Health and Disease
- Vol. 1, 15
Abstract
Background: We previously reported a reduction in central venous catheter (CVC) malfunction when using once-weekly recombinant tissue-plasminogen activator (rt-PA) as a locking solution, compared with thrice-weekly heparin. Objectives: To identify risk factors for CVC malfunction to inform a targeted strategy for rt-PA use. Design: Retrospective analysis. Setting: Canadian hemodialysis (HD) units. Patients: Adults with newly placed tunnelled upper venous system CVCs randomized to a locking solution of rt-PA (1 mg/mL) mid-week and heparin (5000 u/ml) on the other HD sessions, or thrice-weekly heparin (5000 u/ml). Measurements: CVC malfunction (the primary outcome) was defined as: peak blood flow less than 200 mL/min for thirty minutes during a HD session; mean blood flow less than 250 mL/min for two consecutive HD sessions; inability to initiate HD. Methods: Cox regression was used to determine the association between patient demographics, HD session CVC-related variables and the outcome of CVC malfunction. Results: Patient age (62.4 vs 65.4 yr), proportion female sex (35.6 vs 48.4%), and proportion with first catheter ever (60.7 vs 61.3%) were similar between patients with and without CVC malfunction. After multivariate analysis, risk factors for CVC malfunction were mean blood processed < 65 L when compared with ≥ 85 L in the prior 6 HD sessions (HR 4.36; 95% CI, 1.59 to 11.95), and mean blood flow < 300 mL/min, or 300 – 324 mL/min in the prior 6 HD sessions (HR 7.65; 95% CI, 2.78 to 21.01, and HR 5.52; 95% CI, 2.00 to 15.23, respectively) when compared to ≥ 350 mL/min. Limitations: This pre-specified post-hoc analysis used a definition of CVC malfunction that included blood flow, which may result in an overestimate of the effect size. Generalizability of results to HD units where trisodium citrate locking solution is used may also be limited. Conclusions: HD session characteristics including mean blood processed and mean blood flow were associated with CVC malfunction, while patient characteristics were not. Whether targeting these patients at greater risk of CVC malfunction with rt-PA as a locking solution improves CVC longevity remains to be determined.Keywords
This publication has 28 references indexed in Scilit:
- Patterns of Hemodialysis Catheter Dysfunction Defined According to National Kidney Foundation Guidelines As Blood Flow <300 mL/minInternational Journal of Nephrology, 2011
- Does the Heparin Lock Concentration Affect Hemodialysis Catheter Patency?Clinical Journal of the American Society of Nephrology, 2010
- Concentration of Heparin-Locking Solution and Risk of Central Venous Hemodialysis Catheter MalfunctionASAIO Journal, 2007
- Relationship between Blood Flow in Central Venous Catheters and Hemodialysis AdequacyClinical Journal of the American Society of Nephrology, 2006
- IntroductionJournal of the American Society of Nephrology, 2006
- Optimizing Dialysis Delivery in Tunneled Dialysis CathetersASAIO Journal, 2006
- Short and long alteplase dwells in dysfunctional hemodialysis cathetersHemodialysis International, 2005
- Interventional Nephrology and Dialysis: Managing the Complications of Long‐Term Tunneled Dialysis CathetersSeminars in Dialysis, 2003
- A prospective study of complications associated with cuffed, tunnelled haemodialysis catheters†Nephrology Dialysis Transplantation, 2001
- Type of vascular access and mortality in U.S. hemodialysis patientsKidney International, 2001