Dose Performance of a 64-Channel Dual-Source CT Scanner1

Authors:
  • APAndrew N. Primak
    Andrew N. Primak
  • OSOsama Saba
    Osama Saba
  • HBHerbert Bruder
    Herbert Bruder
  • KSKarl Stierstorfer
    Karl Stierstorfer
  • RRRainer Raupach
    Rainer Raupach
  • CSChristoph Suess
    Christoph Suess
  • BSBernhard Schmidt
    Bernhard Schmidt
  • BOBernd M. Ohnesorge
    Bernd M. Ohnesorge
  • TFThomas G. Flohr
    Thomas G. Flohr

Abstract
Purpose: To prospectively compare the dose performance of a 64-channel multi–detector row computed tomographic (CT) scanner and a 64-channel dual-source CT scanner from the same manufacturer. Materials and Methods: To minimize dose in the cardiac (dual-source) mode, the evaluated dual-source CT system uses a cardiac beam-shaping filter, three-dimensional adaptive noise reduction, heart rate–dependent pitch, and electrocardiographically based modulation of the tube current. Weighted CT dose index per 100 mAs was measured for the head, body, and cardiac beam-shaping filters. Kerma-length product was measured in the spiral cardiac mode at four pitch values and three electrocardiographic modulation temporal windows. Noise was measured in an anthropomorphic phantom. Data were compared with data from a 64-channel multi–detector row CT scanner. Results: For the multi–detector row and dual-source CT systems, respectively, weighted CT dose index per 100 mAs was 14.2 and 12.2 mGy (head CT), 6.8 and 6.4 mGy (body CT...