Hip joint center lateralization minimally affects the biomechanics of patient‐specific flanged acetabular components: A computational model
- 26 April 2024
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 42 (10) , 2228-2236
Abstract
Patient‐specific flanged acetabular components are utilized to treat failed total hip arthroplasties with large acetabular defects. Previous clinical studies from our institution showed that these implants tend to lateralize the acetabular center of rotation. However, the clinical impact of lateralization on implant survivorship is debated. Our goal was to develop a finite element model to quantify how lateralization of the native hip center affects periprosthetic strain and implant‐bone micromotion distributions in a static level gait loading condition. To build the model, we computationally created a superomedial acetabular defect in a computed tomography 3D reconstruction of a native pelvis and designed a flanged acetabular implant to address this simulated bone defect. We modeled two implants, one with ~1 cm and a second with ~2 cm of hip center lateralization. We applied the maximum hip contact force and corresponding abductor force observed during level gait. The resulting strains were compared to bone fatigue strength (0.3% strain) and the micromotions were compared to the threshold for bone ingrowth (20 µm). Overall, the model demonstrated that the additional lateralization only slightly increased the area of bone at risk of failure and decreased the areas compatible with bone ingrowth. This computational study of patient‐specific acetabular implants establishes the utility of our modeling approach. Further refinement will yield a model that can explore a multitude of variables and could be used to develop a biomechanically‐based acetabular bone loss classification system to guide the development of patient‐specific implants in the treatment of large acetabular bone defects.Keywords
Funding Information
- Stavros Niarchos Foundation
This publication has 17 references indexed in Scilit:
- The role of muscle forces and gait cycle discretization when assessing acetabular cup primary stability: A finite element studyComputer Methods and Programs in Biomedicine, 2023
- Three-Dimensional Functional Impingement in Total Hip Arthroplasty: A Biomechanical AnalysisThe Journal of Arthroplasty, 2022
- A finite element analysis of relationship between fracture, implant and tibial tunnelScientific Reports, 2021
- Survivorship and Clinical Outcomes of Custom Triflange Acetabular Components in Revision Total Hip Arthroplasty: A Systematic ReviewThe Journal of Arthroplasty, 2019
- Mechanical performance of cementless total knee replacements: It is not all about the maximum loadsJournal of Orthopaedic Research, 2018
- Finite Element Analysis of the Pelvis after Customized Prosthesis ReconstructionJournal of Bionic Engineering, 2018
- Midterm Survivorship After Revision Total Hip Arthroplasty With a Custom Triflange Acetabular ComponentThe Journal of Arthroplasty, 2018
- Outcomes of Custom Flange Acetabular Components in Revision Total Hip Arthroplasty and Predictors of FailureThe Journal of Arthroplasty, 2016
- Computational assessment of the effect of polyethylene wear rate, mantle thickness, and porosity on the mechanical failure of the acetabular cement mantleJournal of Orthopaedic Research, 2009
- Anisotropy of the fatigue behaviour of cancellous boneJournal of Biomechanics, 2008