MRI and implantable cardiac electronic devices
Authors:
- Sebastian JäckleSJSebastian Jäckle
- Vlada ZverevaVZVlada Zvereva
- 1 January 2015
- journal article
- research article
- Published by Wolters Kluwer Health in Current Opinion in Cardiology
- Vol. 30 (1) , 65-73
Abstract
To assess the current knowledge about the potential hazard from MRI in patients with devices such as pacemakers and implantable cardioverter defibrillators (ICDs).Most data concern 'MRI unsafe' devices, with only a few studies on 'MRI conditional' devices. No 'MRI safe' cardiac devices are currently available. Studies on 'MRI unsafe' devices tend to be small scale and reflect the experience of individual centres; few provide long-term follow-up data. Many newer devices are approved as 'MRI conditional' based on technical simulations or postmarket surveillance studies. With adequate measures taken before performing an MRI scan, reported complication rates are generally low, but there is a nonnegligible residual risk for power-on reset and lead heating. The presence of abandoned, older leads may affect the propensity for lead heating during MRI with newer devices, including those designated 'MRI conditional'. Very little research has been carried out on the hazard from MRI scans in patients with ICDs, but registry data indicate more events with ICDs than with pacemakers.The limited available data indicate a manageable but not negligible MRI-associated hazard in patients with implantable cardiac devices. Further controlled studies and large, independent registries, particularly in Europe, are needed to provide important safety information.Keywords
This publication has 46 references indexed in Scilit:
- Determining the risks of clinically indicated nonthoracic magnetic resonance imaging at 1.5 T for patients with pacemakers and implantable cardioverter-defibrillators: Rationale and design of the MagnaSafe RegistryAmerican Heart Journal, 2013
- ACR guidance document on MR safe practices: 2013Journal of Magnetic Resonance Imaging, 2013
- Safe magnetic resonance image scanning of the pacemaker patient: current technologies and future directionsEP Europace, 2012
- Monocenter feasibility study of the MRI compatibility of the Evia pacemaker in combination with Safio S pacemaker leadJournal of Cardiovascular Magnetic Resonance, 2012
- Initial experience with magnetic resonance imaging-safe pacemakersJournal of Interventional Cardiac Electrophysiology, 2011
- Safety of serial MRI in patients with implantable cardioverter defibrillatorsHeart, 2011
- Magnetic Resonance Imaging of Implantable Cardiac Rhythm Devices at 3.0 TeslaPacing and Clinical Electrophysiology, 2008
- Clinical Utility and Safety of a Protocol for Noncardiac and Cardiac Magnetic Resonance Imaging of Patients With Permanent Pacemakers and Implantable-Cardioverter Defibrillators at 1.5 TeslaCirculation, 2006
- In vivo heating of pacemaker leads during magnetic resonance imagingEuropean Heart Journal, 2004
- Effects of magnetic resonance imaging on cardiac pacemakers and electrodesAmerican Heart Journal, 1997