Effect of pulse phase duration on forward masking and spread of excitation in cochlear implant listeners
Open Access
- 20 July 2020
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 15 (7) , e0236179
Abstract
Previous cochlear implant (CI) research has shown that at a pulse train with a long pulse phase duration (PPD) requires less current but greater charge to obtain the same loudness as a pulse train with a short PPD. This might result in different excitation patterns between long and short PPDs. At equal loudness, long PPDs might produce greater masking due to greater charge. However, because they require less current, long PPDs may produce a smaller spatial spread of excitation (SOE) compared to short PPDs by evoking a greater neural firing probability within the relatively small current field. To investigate the effects of PPD on excitation patterns, overall masking and SOE were compared for equally loud stimuli with short or long PPD in 10 adult CI ears. Forward masking patterns were measured at relatively soft, medium, and loud presentation levels. Threshold shifts were calculated in terms of percent dynamic range (DR) of the probe. The area under the curve (AUC) of the masking functions was significantly larger for the long PPD than for the short PPD masker. The difference in AUC was proportional to the difference in charge between the short and long PPD maskers. To estimate SOE, the masking patterns were first normalized to the peak masking, and then AUC was calculated. SOE was significantly larger for the short PPD than for the long PPD masker. Thus, at equal loudness, long PPDs produced greater overall masking (possibly due to greater charge) but less SOE (possibly due to less current spread) than did short PPDs. The effect of the interaction between masking and SOE by long PPD stimulation remains to be tested.Keywords
Funding Information
- National Institute on Deafness and Other Communication Disorders (R01 DC017702-01)
This publication has 40 references indexed in Scilit:
- Place specificity of monopolar and tripolar stimuli in cochlear implants: The influence of residual maskingThe Journal of the Acoustical Society of America, 2013
- Cochlear-implant spatial selectivity with monopolar, bipolar and tripolar stimulationHearing Research, 2012
- Current focusing sharpens local peaks of excitation in cochlear implant stimulationHearing Research, 2010
- Neural Masking by Sub-threshold Electric Stimuli: Animal and Computer Model ResultsJournal of the Association for Research in Otolaryngology, 2010
- Identifying Cochlear Implant Channels with Poor Electrode-Neuron Interface: Partial Tripolar, Single-Channel Thresholds and Psychophysical Tuning CurvesEar & Hearing, 2010
- Forward-masking patterns produced by symmetric and asymmetric pulse shapes in electric hearingThe Journal of the Acoustical Society of America, 2010
- Forward-masked spatial tuning curves in cochlear implant usersThe Journal of the Acoustical Society of America, 2008
- Changes Across Time in the Temporal Responses of Auditory Nerve Fibers Stimulated by Electric Pulse TrainsJournal of the Association for Research in Otolaryngology, 2008
- Changes Across Time in Spike Rate and Spike Amplitude of Auditory Nerve Fibers Stimulated by Electric Pulse TrainsJournal of the Association for Research in Otolaryngology, 2007
- Elemente der PsychophysikThe American Journal of Psychology, 1889