Dosimetric effect of six degrees of freedom couch top with rotational setup error corrections in proton therapy
Open Access
- 30 May 2023
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 24 (9) , e14043
Abstract
Purpose: To investigate the dosimetric effect of six degrees of freedom (6DoF) couch top with rotational corrections in proton therapy (PT).Methods: The water equivalent thickness (WET) was measured using a proton beam with a 6DoF couch top and patient immobilization base plate (PIBP) placed in front of a motorized water phantom. The accuracy verification was performed with the beam axis set perpendicular to the 6DoF couch top and tilted in 10° steps from 10° to 30°. Up to 3° rotational correction may be added during the actual treatment to correct the rotational setup error on our system. The measured and calculated values using the treatment planning system were compared. Additionally, the effect of the 3° difference was evaluated using actual measurements concerning each angle on the proton beam range.Results: The WET of the 6DoF couch top and PIBP were 8.5 ± 0.1 mm and 6.8 ± 0.1 mm, respectively. The calculation and the actual measurement at each angle agreed within 0.2 mm at the maximum. A maximum difference of approximately 0.6 mm was confirmed when tilted at 3° following 30° with the 6DoF couch top plus PIBP.Conclusions: The dosimetric effect of the 6DoF couch top with rotational corrections in PT differs depending on the incidence angle on the couch top, and it increased with the increased oblique angle of incidence. However, the effect on the range was as small as 0.6 mm at the maximum. The amount of rotational correction, the angle of incidence of the beam, and the effect of rotational corrections on the proton beam range may differ depending on the structure of the couch top. Therefore, sufficient prior confirmation, and subsequent periodical quality assurance management are important.Keywords
This publication has 13 references indexed in Scilit:
- A model-based analysis of a simplified beam-specific dose output in proton therapy with a single-ring wobbling systemPhysics in Medicine & Biology, 2014
- Immobilization Considerations for Proton Radiation TherapyTechnology in Cancer Research & Treatment, 2014
- Dosimetric effects caused by couch tops and immobilization devices: Report of AAPM Task Group 176Medical Physics, 2014
- Experimental verification of dose calculation using the simplified Monte Carlo method with an improved initial beam model for a beam-wobbling systemPhysics in Medicine & Biology, 2013
- Water Equivalent Thickness Analysis of Immobilization Devices for Clinical Implementation in Proton TherapyTCRT Express, 2013
- Dosimetric characterization of the iBEAM evo carbon fiber couch for radiotherapyMedical Physics, 2010
- Dose perturbations by two carbon fiber treatment couches and the ability of a commercial treatment planning system to predict these effectsMedical Physics, 2009
- Modeling of a digital couch for a proton treatment planning systemJournal of the Korean Physical Society, 2009
- Modeling of carbon fiber couch attenuation properties with a commercial treatment planning systemMedical Physics, 2008
- The effect of gantry angle on megavoltage photon beam attenuation by a carbon fiber couch insertMedical Physics, 2005