Efficient on‐line setup correction strategies using plan‐intent functions
Authors:
- Harry KellerHKHarry Keller
- David A. JaffrayDJDavid A. Jaffray
- Tara RosewallTRTara Rosewall
- Elizabeth WhiteEWElizabeth White
- 27 April 2006
- journal article
- Published by Wiley in Medical Physics
- Vol. 33 (5) , 1388-1397
Abstract
With the introduction of image-guided radiation therapy (IGRT) delivery systems on-line set-up correction strategies have gained in popularity. Usually, the correction workload of these strategies is high compared to off-line strategies as daily setup corrections have to be performed based on a predefined action level. In this work, it is proposed that on-line strategies must not only be judged in terms of workload but also in terms of efficacy. While workload can be easily predicted for such strategies, the efficacy must ultimately reflect the efficiency with which the original treatment plan intent is met. The purpose of this work is to investigate the tradeoff between workload and efficacy of three different on-line set-up correction strategies: The common fixed action level strategy and two novel on-line setup correction strategies, i.e., a dose-volume histogram (DVH) constraint and an equivalent uniform dose (EUD) score strategy that aim directly for better compliance with original treatment plan intent. All strategies were reformulated in terms of a score function that reflected treatment plan intent. A retrospective study was conducted on 5 prostate patients (7-field conformal, 79.8 Gy, 42 fractions). PTV margins were 10 mm except in the posterior direction (7 mm). The original treatment plan intent for these patients was defined using a set of DVH constraints. The results show that the on-line setup correction strategy based on a fixed action level of 3 mm resulted in a considerable correction workload. For larger action levels, a dose benefit (in terms of EUD) in the rectum and bladder was observed for all patients which is clinically "fortuitous" but difficult to take advantage of. In contrast, the application of the two novel strategies generally resulted in a controlled decrease of the dose to the rectum and the bladder with a smaller workload. It is concluded that using information about target anatomy and the planned dose distribution allows the design of specific correction strategies that are better tailored to the individual patient and that comply effectively with initial treatment plan intent.Keywords
This publication has 19 references indexed in Scilit:
- Prostate gland motion assessed with cine-magnetic resonance imaging (cine-MRI)International Journal of Radiation Oncology - Biology - Physics (IJROBP), 2005
- On-line aSi portal imaging of implanted fiducial markers for the reduction of interfraction error during conformal radiotherapy of prostate carcinomaInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2004
- An integrated volumetric imaging and guidance system for targeting of soft-tissue structures in radiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2003
- Inverse planning incorporating organ motionMedical Physics, 2000
- Reporting and analyzing dose distributions: A concept of equivalent uniform doseMedical Physics, 1997
- Adaptive radiation therapyPhysics in Medicine & Biology, 1997
- Combined error of patient positioning variability and prostate motion uncertainty in 3D conformal radiotherapy of localized prostate cancerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1996
- High-precision prostate cancer irradiation by clinical application of an offline patient setup verification procedure, using portal imagingInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1996
- A new model for “accept or reject” strategies in off-line and on-line megavoltage treatment evaluationInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1995
- A verification procedure to improve patient set-up accuracy using portal imagesRadiotherapy and Oncology, 1993