Empiric Validation of Simulation Models for Estimating Glucose Meter Performance Criteria for Moderate Levels of Glycemic Control
Authors:
- 1 December 2013
- journal article
- research article
- Published by SAGE Publications in Diabetes Technology & Therapeutics
- Vol. 15 (12) , 996-1003
Abstract
Background: We used simulation modeling to relate glucose meter performance criteria to insulin dosing errors for patients on a moderate glycemic control protocol (glucose target, 110–150 mg/dL) and empirically validated assumptions from simulation models using observed glucose meter and laboratory glucose values obtained nearly simultaneously. Subjects and Methods: The 25,948 glucose values from 1,513 patients on a moderate glycemic control protocol were used to represent the expected distribution of glucose values in this patient population. Simulation models were used to relate glucose meter analytical performance to insulin dosing errors assuming 10%, 15%, or 20% total allowable error (TEa). In addition, 4,017 paired glucose meter and serum laboratory glucose measurements drawn within 5 min of each other were used to generate an empiric dataset to validate simulation model assumptions relating glucose meter performance to insulin dosing errors. Results: Large (three or more category) insulin dosing errors are predicted to occur only under the 20% TEa condition. Two category insulin dosing errors were common (6–20% of all insulin dosing decisions) when 20% TEa was assumed, but frequency decreased to only 0.2% of dosing decisions when 10% TEa was modeled. When insulin dosing error rates were measured empirically by comparing paired glucose meter and laboratory glucose values, insulin dosing error rates were very similar to those predicted for the 20% TEa condition. Conclusions: Both simulation models and empiric data demonstrate that glucose meters that perform at ≥20% TEa allow large insulin dosing errors during a moderate glycemic control protocol.Keywords
This publication has 19 references indexed in Scilit:
- Simulating Total Error While Excluding Results Underestimates Total ErrorClinical Chemistry, 2010
- Glucose Meter Performance Criteria for Tight Glycemic Control Estimated by Simulation ModelingClinical Chemistry, 2010
- Intensive versus Conventional Glucose Control in Critically Ill PatientsNew England Journal of Medicine, 2009
- Chapter 16 Monte Carlo Simulation in Establishing Analytical Quality Requirements for Clinical Laboratory Tests2009
- Tight Glucose Control in the Intensive Care Unit: Are Glucose Meters up to the Task?Clinical Chemistry, 2009
- Evaluation of the Impact of Hematocrit and Other Interference on the Accuracy of Hospital-Based Glucose MetersDiabetes Technology & Therapeutics, 2008
- Evaluation of point-of-care glucose testing accuracy using locally-smoothed median absolute difference curvesClinica Chimica Acta, 2008
- Accuracy of Roche Accu-Chek Inform Whole Blood Capillary, Arterial, and Venous Glucose Values in Patients Receiving Intensive Intravenous Insulin Therapy After Cardiac SurgeryAmerican Journal of Clinical Pathology, 2007
- Glucose Measurement: Confounding Issues in Setting Targets for Inpatient ManagementDiabetes Care, 2007
- Intensive Insulin Therapy in Critically Ill PatientsNew England Journal of Medicine, 2001