Glucose Meter Performance Criteria for Tight Glycemic Control Estimated by Simulation Modeling
Authors:
Open Access
- 1 July 2010
- journal article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 56 (7) , 1091-1097
Abstract
Background: Glucose meter analytical performance criteria required for safe and effective management of patients on tight glycemic control (TGC) are not currently defined. We used simulation modeling to relate glucose meter performance characteristics to insulin dosing errors during TGC.Methods: We used 29 920 glucose values from patients on TGC at 1 institution to represent the expected distribution of glucose values during TGC, and we used 2 different simulation models to relate glucose meter analytical performance to insulin dosing error using these 29 920 initial glucose values and assuming 10%, 15%, or 20% total allowable error (TEa) criteria.Results: One-category insulin dosing errors were common under all error conditions. Two-category insulin dosing errors occurred more frequently when either 20% or 15% TEa was assumed compared with 10% total error. Dosing errors of 3 or more categories, those most likely to result in hypoglycemia and thus patient harm, occurred infrequently under all error conditions with the exception of 20% TEa.Conclusions: Glucose meter technologies that operate within a 15% total allowable error tolerance are unlikely to produce large (≥3-category) insulin dosing errors during TGC. Increasing performance to 10% TEa should reduce the frequency of 2-category insulin dosing errors, although additional studies are necessary to determine the clinical impact of such errors during TGC. Current criteria that allow 20% total allowable error in glucose meters may not be optimal for patient management during TGC.Keywords
This publication has 21 references indexed in Scilit:
- Intensive versus Conventional Glucose Control in Critically Ill PatientsNew England Journal of Medicine, 2009
- Chapter 16 Monte Carlo Simulation in Establishing Analytical Quality Requirements for Clinical Laboratory Tests2009
- Accuracy of bedside glucose measurement from three glucometers in critically ill patients*Critical Care Medicine, 2008
- Comparison of POCT and central laboratory blood glucose results using arterial, capillary, and venous samples from MICU patients on a tight glycemic protocolClinica Chimica Acta, 2008
- Accuracy of Glycemic Measurements in the Critically IllDiabetes Technology & Therapeutics, 2008
- Evaluation of point-of-care glucose testing accuracy using locally-smoothed median absolute difference curvesClinica Chimica Acta, 2008
- Severe hypoglycemia in critically ill patients: Risk factors and outcomes*Critical Care Medicine, 2007
- Accuracy of Roche Accu-Chek Inform Whole Blood Capillary, Arterial, and Venous Glucose Values in Patients Receiving Intensive Intravenous Insulin Therapy After Cardiac SurgeryAmerican Journal of Clinical Pathology, 2007
- Glucose Measurement: Confounding Issues in Setting Targets for Inpatient ManagementDiabetes Care, 2007
- Intensive Insulin Therapy in Critically Ill PatientsNew England Journal of Medicine, 2001