4D Flow MRI ‐Based Quantification of Tricuspid Regurgitation: A Possible Approach to Improve Measurement Repeatability
Authors:
- Alessandra RivaARAlessandra Riva
- Giulia PasqualinGPGiulia Pasqualin
- Francesco SturlaFSFrancesco Sturla
- Lara TondiLTLara Tondi
- Simone SaittaSSSimone Saitta
- Davide TondiDTDavide Tondi
- Jessica ZannoniJZJessica Zannoni
- Giandomenico DisabatoGDGiandomenico Disabato
- Daniel GieseDGDaniel Giese
- Massimo ChessaMCMassimo Chessa
- Pietro SpagnoloPSPietro Spagnolo
- Alberto RedaelliARAlberto Redaelli
- 9 July 2026
- journal article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 96 (5) , 2381-2395
Abstract
Purpose: To address limitations in tricuspid regurgitation (TR) quantification by magnetic resonance imaging (MRI) or transthoracic echocardiography (TTE), a new method is proposed to directly quantify TR volume (TR V ) using 4D Flow MRI with retrospective tricuspid valve (TV) annular tracking. Methods: Thirty patients with TR were prospectively enrolled. TR V was quantified using four MRI‐based methods: (i) indirect, (ii) direct (2D phase‐contrast), (iii) 4D Flow with a static TV plane, and (iv) 4D Flow with dynamic annular tracking. Reproducibility and agreement were assessed in terms of TR grading against standard MRI‐ and TTE‐based methods. Results: The dynamic 4D Flow method correlated strongly with all other MRI methods ( r = 0.77–0.81, p < 0.0001) but consistently yielded higher TR V values. Compared with the indirect method, it showed a Bland–Altman bias of 16 mL (limits of agreement: −19 to 51 mL), along with the highest agreement (intraclass correlation coefficient of 0.88). Agreement in TR severity grading was poor across methods (kappa ≤ 0.07), with dynamic 4D Flow classifying more cases as moderate or severe. Agreement with TTE was slight (kappa = 0.17, p = 0.27) with concordant TR grading in 50% of patients. Inter‐observer reproducibility was excellent ( r = 0.99, bias = 0.7 mL, kappa = 0.80). Conclusion: 4D Flow with dynamic TV tracking offers a reproducible and potentially more sensitive method for TR V quantification than conventional MRI techniques. Despite limited alignment with TTE, this technique may help overcome key imaging limitations and improve TR clinical evaluation, especially in complex cases.Keywords
Funding Information
- Ministero della Salute
This publication has 34 references indexed in Scilit:
- Prognostic Implications of Functional Tricuspid Regurgitation in Asymptomatic Degenerative Mitral RegurgitationESC Heart Failure, 2025
- Tricuspid valve flow measurement using a deep learning framework for automated valve‐tracking 2D phase contrastMagnetic Resonance in Medicine, 2024
- Mitral valve regurgitation assessed by intraventricular CMR 4D-flow: a systematic review on the technological aspects and potential clinical applicationsThe International Journal of Cardiovascular Imaging, 2023
- Comparison of Four‐Dimensional Magnetic Resonance Imaging Analysis of Left Ventricular Fluid Dynamics and Energetics in Ischemic and Restrictive CardiomyopathiesJournal of Magnetic Resonance Imaging, 2022
- 4D flow MRI quantification of mitral and tricuspid regurgitation: Reproducibility and consistency relative to conventional MRIJournal of Magnetic Resonance Imaging, 2018
- Pressure overloaded right ventricles: a multicenter study on the importance of trabeculae in RV function measured by CMRThe International Journal of Cardiovascular Imaging, 2014
- MR velocity mapping of tricuspid flow: Correction for through‐plane motionJournal of Magnetic Resonance Imaging, 1997
- Temporal phase unwrapping for cine velocity imagingJournal of Magnetic Resonance Imaging, 1995
- Analysis of systematic and random error in MR volumetric flow measurementsMagnetic Resonance in Medicine, 1993
- Semiautomated method for noise reduction and background phase error correction in MR phase velocity dataJournal of Magnetic Resonance Imaging, 1993