The influence of different technique factors on image quality of lumbar spine radiographs as evaluated by established CEC image criteria.
Authors:
- A TingbergATA Tingberg
- S MattssonSMS Mattsson
- J BesjakovJBJ Besjakov
- S KheddacheSKS Kheddache
- B LanhedeBLB Lanhede
- L G MånssonLML G Månsson
- M ZanklMZM Zankl
- 1 November 2000
- journal article
- research article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 73 (875) , 1192-1199
Abstract
In this study we have investigated the image quality of lumbar spine radiographs taken after recording technical and physical parameters. Two technical parameters were altered, tube voltage (70 kV and 90 kV for the anteroposterior (AP) projection and 77 kV and 95 kV for the lateral projection) and sensitivity of the film-screen system (sensitivity class 400 and 600). In total, 85 images were included in the study. Entrance surface dose (ESD) was measured using thermoluminescent dosemeters. The mean value of ESD for the different technique groups varied between 1.9 mGy (90 kV, sensitivity class 400) and 4.6 mGy (70 kV, sensitivity class 400) for the AP projection, and between 6.4 mGy (95 kV, sensitivity class 600) and 20.4 mGy (70 kV, sensitivity class 400) for the lateral projection. Image criteria given in the "European Guidelines on Quality Criteria for Radiographic Images" were used to assess image quality. Two evaluation methods have been employed. A straightforward scoring of fulfilled image criteria, and visual grading analysis using the structures defined in the image criteria. The latter method provided a sharper distinction between groups of images taken using different radiographic techniques. The average number of fulfilled image criteria for the AP projections varied between 0.74 (90 kV, sensitivity class 400) and 0.87 (70 kV, sensitivity class 400). For the lateral projection this number varied between 0.79 (95 kV, sensitivity class 600) and 0.84 (77 kV, sensitivity class 600). This study shows that image criteria are useful tools in clinical studies of image quality.Keywords
This publication has 14 references indexed in Scilit:
- The use of carbon fibre material in radiographic cassettes: estimation of the dose and contrast advantages.British Journal of Radiology, 1997
- Examination Technique, Image Quality, and Patient Dose in Paediatric RadiologyActa Radiologica, 1996
- Evaluation of the European image quality criteria for chest examinationsBritish Journal of Radiology, 1995
- Optimization of radiographic technique for chest radiographyBritish Journal of Radiology, 1995
- CEC Quality Criteria for Diagnostic Radiographic Images - Basic ConceptsRadiation Protection Dosimetry, 1995
- Monte Carlo study of grid performance in diagnostic radiology: task dependent optimization for screen–film imagingBritish Journal of Radiology, 1994
- In search of optimum chest radiography techniquesBritish Journal of Radiology, 1993
- Doses to Patients From Diagnostic Radiology in FranceHealth Physics, 1988
- Patient doses and risks from diagnostic radiology in North-east ItalyBritish Journal of Radiology, 1987
- Doses to patients from routine diagnostic X-ray examinations in EnglandBritish Journal of Radiology, 1986