Abstract
Clinical microbeam radiation therapy (MRT) will require a multislit collimator with adjustable uniform slit widths to enable reliable Monte Carlo-based treatment planning. Such a collimator has been designed, fabricated of tungsten [W] by Tecomet/Viasys (Woburn, Massachusetts, USA) and installed at the electron-wiggler-generated hard x-ray ID17 beamline of the European Synchrotron Radiation Facility. Its pair of 125 parallel, deep, wide radiolucent slits, on center, are perfused with nitrogen gas to dissipate heat during irradiation. Major improvements in uniformity of microbeam widths and good peak/valley dose ratios combined with a very high dose rate in targeted tissues have been achieved.
Keywords
This publication has 13 references indexed in Scilit:
- MOSFET dosimetry for microbeam radiation therapy at the European Synchrotron Radiation FacilityMedical Physics, 2003
- High sensitivity radiochromic film dose comparisonsPhysics in Medicine & Biology, 2002
- Weanling piglet cerebellum: a surrogate for tolerance to MRT (microbeam radiation therapy) in pediatric neuro-oncology2001
- Monte Carlo simulation of dose distributions from a synchrotron-produced microplanar beam array using the EGS4 code system4Physics in Medicine & Biology, 2000
- Physics study of microbeam radiation therapy with PSI‐version of Monte Carlo code GEANT as a new computational toolMedical Physics, 2000
- XOP: recent developments1998
- XOP: a multiplatform graphical user interface for synchrotron radiation spectral and optics calculations1997
- Subacute neuropathological effects of microplanar beams of x-rays from a synchrotron wiggler.Proceedings of the National Academy of Sciences, 1995
- Microbeam radiation therapyMedical Physics, 1992
- Photon energy dependence of the sensitivity of radiochromic film and comparison with silver halide film and LiF TLDs used for brachytherapy dosimetryMedical Physics, 1991