Anterior chamber depth measurements using Scheimpflug imaging and optical coherence tomography: Repeatability, reproducibility, and agreement
- 1 January 2015
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 41 (1) , 178-185
Abstract
To determine the repeatability, reproducibility, and agreement of anterior chamber depth (ACD) measurements obtained with 3 Scheimpflug cameras and an anterior segment optical coherence tomography (AS-OCT) device. Eye Hospital of Wenzhou Medical University, Zhejiang, China. Observational cross-sectional study. Two observers took 3 consecutive measurements in healthy right eyes using each device to assess intraoperator repeatability. The mean values obtained at different sessions by the first operator were used to determine the intersession reproducibility. Three consecutive measurements obtained by the first operator at the first session were averaged and used to assess agreement. The ACD measurements obtained by 2 observers in 71 eyes were highly repeatable using the 4 devices, with a test-retest repeatability of 0.04 to 0.07 mm for intraoperator repeatability. The interoperator and intersession reproducibility of ACD measurements were high, and the test-retest repeatability ranges of interoperator and intersession reproducibility were 0.06 to 0.07 mm and 0.05 to 0.08 mm, respectively. The ACD measurements of the 4 systems were sorted from the thickest to the thinnest (Galilei G2 > Visante > Sirius > Pentacam). The differences in the measurements were statistically significant except between the ACD measurements obtained by the Sirius device and the Visante device. However, good agreement with narrow 95% limits of agreement was found between these devices. The 4 devices provided high intraoperator repeatability and interoperator and intersession reproducibility for ACD measurements. Good agreement in ACD measurements was found between the devices in healthy eyes. No author has a financial or proprietary interest in any material or method mentioned.Keywords
This publication has 24 references indexed in Scilit:
- Comparison of anterior chamber depth measurements by 3-dimensional optical coherence tomography, partial coherence interferometry biometry, Scheimpflug rotating camera imaging, and ultrasound biomicroscopyJournal of Cataract & Refractive Surgery, 2012
- Definition of ACDOphthalmology, 2011
- Anterior Chamber Depth Measurement in Pseudophakic Eyes: A Comparison of Pentacam and UltrasoundJournal of Refractive Surgery, 2010
- Comparison of partial coherence interferometry and ultrasound for anterior segment biometryJournal of Cataract & Refractive Surgery, 2009
- Meridional analysis for calculating the expected spherocylindrical refraction in eyes with toric intraocular lensesJournal of Cataract & Refractive Surgery, 2007
- Posterior chamber phakic intraocular lens (ICL) for hyperopia: Ten-year follow-upJournal of Cataract & Refractive Surgery, 2007
- Prediction of the effective postoperative (intraocular lens) anterior chamber depthJournal of Cataract & Refractive Surgery, 2006
- Comparison of Three Methods of Measuring Corneal Thickness and Anterior Chamber DepthAmerican Journal of Ophthalmology, 2006
- Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to HaigisGraefe's Archive for Clinical and Experimental Ophthalmology, 2000
- Ultrasound biomicroscopic and conventional ultrasonographic study of ocular dimensions in primary angle-closure glaucoma11Conducted on behalf of the Regional Council of the Veneto Region, Finalized Health Research, Venice, Italy.Ophthalmology, 1998