Formulation and selection of composite biomaterials for acetabular cup of hip prosthesis with wear analysis
- 13 June 2023
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part E: Journal of Process Mechanical Engineering
- Vol. 238 (6) , 2983-2993
Abstract
The study of tribological performance of metal-on-plastic type of hip prostheses has drawn the attention of researchers since the last few decades. These studies mainly focus on the friction and wear behaviour of acetabular cups made of different polymer biomaterials with a view to enhance the prostheses life. In this study, a novel biomaterial coded as LHHA has been prepared for the acetabular cup in three different blends with help of bioprocess engineering by mixing low-density polyethylene and high-density polyethylene with10, 20 and 30 percent of hydroxyapatite to improve the mechanical properties. A popular decision-making method was applied to select the best alternative from these three blends on the basis of hardness, roughness and ultimate strength and LHHA-30 is found to be the best. Furthermore, experimental wear investigation was carried out in a well-designed hip simulator to control such biological system under a series of random loads, within 600N and 1500N acting at different points on prosthesis for 10 million gait cycles run followed by theoretical wear analysis. Wear rates are found to be 12.651 and 26.247 mg per million cycle with marginal differences of 0.1 and 0.6 for LHHA-30 and ultrahigh molecular weight polyethylene (UHMWPE) acetabular cups against cobalt–chromium (Co–Cr) alloy femoral head, respectively. Morphological studies carried out by scanning electron microscope revealed that LHHA-30 scored less compared to UHMWPE surface.Keywords
Funding Information
- All India Council for Technical Education (15EDMER005)
- Ministry of Human Resource Development (15EDMER005)
This publication has 18 references indexed in Scilit:
- Effect of contact area on the wear of ultrahigh molecular weight polyethylene in noncyclic pin-on-disk testsTribology International, 2017
- In Vivo Damage of the Head-Neck Junction in Hard-on-Hard Total Hip Replacements: Effect of Femoral Head Size, Metal Combination, and 12/14 Taper DesignMaterials, 2017
- Importância da avaliação pré‐clínica do desgaste em projetos de implantes de quadril usando máquinas simuladorasRevista Brasileira de Ortopedia, 2017
- Standardized Loads Acting in Hip ImplantsPLOS ONE, 2016
- Metal-on-polyethylene versus metal-on-metal bearing surfaces in total hip arthroplastyThe Bone & Joint Journal, 2015
- Investigation of wear and wear particles from a UHMWPE/multi-walled carbon nanotube nanocomposite for total joint replacementsWear, 2014
- Material selection using preferential ranking methodsMaterials & Design (1980-2015), 2012
- Hip Joint Simulators: State of the Art2006
- Biological reactions to wear debris in total joint replacementProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2000