Experimental Validation of Perfusion Imaging With HOSVD Clutter Filters
- 21 April 2020
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 67 (9) , 1830-1838
Abstract
Novel pulsed-Doppler methods for perfusion imaging are validated using dialysis cartridges as perfusion phantoms. Techniques that were demonstrated qualitatively at 24 MHz, in vivo, are here examined quantitatively at 5 and 12.5 MHz using phantoms with the blood-mimicking fluid flow within cellulose microfibers. One goal is to explore a variety of flow states to optimize measurement sensitivity and flow accuracy. The results show that 2-3-s echo acquisitions at roughly 10 frames/s yield the highest sensitivity to flows of 1-4 mL/min. A second goal is to examine methods for setting the parameters of higher order singular value decomposition (HOSVD) clutter filters. For stationary or moving clutter, the velocity of the blood-mimicking fluid in the microfibers is consistently estimated within measurement uncertainty (mean coefficient of variation = 0.26). Power Doppler signals were equivalent for stationary and moving clutter after clutter filtering, increasing approximately 3 dB/mL/min of blood-mimicking fluid flow for $0\leq {q}\leq 4$ mL/min. Comparisons between phantom and preclinical images show that peripheral perfusion imaging can be reliably achieved without contrast enhancement.
Keywords
Funding Information
- National Heart, Lung, and Blood Institute of the National Institutes of Health (R01HL148664)
This publication has 29 references indexed in Scilit:
- Spatiotemporal Clutter Filtering of Ultrafast Ultrasound Data Highly Increases Doppler and fUltrasound SensitivityIEEE Transactions on Medical Imaging, 2015
- Microvascular imaging: techniques and opportunities for clinical physiological measurementsPhysiological Measurement, 2014
- Comparison of global estimates of prevalence and risk factors for peripheral artery disease in 2000 and 2010: a systematic review and analysisThe Lancet, 2013
- A New Truncation Strategy for the Higher-Order Singular Value DecompositionSIAM Journal on Scientific Computing, 2012
- The Singular Value Filter: A General Filter Design Strategy for PCA-Based Signal Separation in Medical Ultrasound ImagingIEEE Transactions on Medical Imaging, 2011
- Real-time adaptive clutter rejection filtering in color flow imaging using power method iterationsIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2006
- Adaptive Clutter Filtering via Blind Source Separation for Two-Dimensional Ultrasonic Blood Velocity MeasurementUltrasonic Imaging, 2002
- Performance of ultrasound echo decomposition using singular spectrum analysisUltrasound in Medicine & Biology, 2001
- A Multilinear Singular Value DecompositionSIAM Journal on Matrix Analysis and Applications, 2000
- Real-Time Two-Dimensional Blood Flow Imaging Using an Autocorrelation Technique1985