Efficiency of Methods for Karl Fischer Determination of Water in Oils Based on Oven Evaporation and Azeotropic Distillation
Authors:
- 7 February 2003
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 75 (6) , 1227-1232
Abstract
The efficiency of azeotropic distillation and oven evaporation techniques for trace determination of water in oils has recently been questioned by the National Institute of Standards and Technology (NIST), on the basis of measurements of the residual water found after the extraction step. The results were obtained by volumetric Karl Fischer (KF) titration in a medium containing a large excess of chloroform (≥65%), a proposed prerequisite to ensure complete release of water from the oil matrix. In this work, the extent of this residual water was studied by means of a direct zero-current potentiometric technique using a KF medium containing more than 80% chloroform, which is well above the concentration recommended by NIST. A procedure is described that makes it possible to correct the results for dilution errors as well as for chemical interference effects caused by the oil matrix. The corrected values were found to be in the range of 0.6−1.5 ppm, which should be compared with the 12−34 ppm (uncorrected values) reported by NIST for the same oils. From this, it is concluded that the volumetric KF method used by NIST gives results that are much too high.Keywords
This publication has 10 references indexed in Scilit:
- Systematic Bias in the Measurement of Water in Oils by Tubular Oven Evaporation and Azeotropic DistillationAnalytical Chemistry, 2001
- Simultaneous Determination of Dissolved Gases and Moisture in Mineral Insulating Oils by Static Headspace Gas Chromatography with Helium Photoionization Pulsed Discharge DetectionAnalytical Chemistry, 2001
- Bias Assessment of Current Technologies Used for the Determination of Low Levels of Moisture in Mineral Oil SamplesAnalytical Chemistry, 2001
- Determination of Water in NIST Reference Material for Mineral OilsAnalytical Chemistry, 2000
- Determination of the Analytical Performance of a Headspace Capillary Gas Chromatographic Technique and Karl Fischer Coulometric Titration by System Calibration Using Oil Samples Containing Known Amounts of MoistureAnalytical Chemistry, 1999
- Source of the Difference between the Measurement of Water in Hydrocarbons As Determined by the Volumetric and Coulometric Karl Fischer MethodsAnalytical Chemistry, 1999
- Potentiometric Determination of Water Using Spent Imidazole-Buffered Karl Fischer ReagentsAnalytical Chemistry, 1998
- Sources of Systematic Bias in the Measurement of Water by the Coulometric and Volumetric Karl Fischer MethodsAnalytical Chemistry, 1997
- Amperometric Measurement of Moisture in Transformer Oil Using Karl Fischer ReagentsAnalytical Chemistry, 1995
- Comparison of high precision coulometric and West-Gaeke methods with the gravimetric method for preparation of standard sulfur dioxide gas blends using permeation tubesAnalytical Chemistry, 1975