Anatomical Model Uncertainty for RF Safety Evaluation of Metallic Implants Under MRI Exposure
- 8 August 2019
- journal article
- research article
- Published by Wiley in Bioelectromagnetics
- Vol. 40 (7) , 458-471
Abstract
The Virtual Population (ViP) phantoms have been used in many dosimetry studies, yet, to date, anatomical phantom uncertainty in radiofrequency (RF) research has largely been neglected. The objective of this study is to gain insight, for the first time, regarding the uncertainty in RF‐induced fields during magnetic resonance imaging associated with tissue assignment and segmentation quality and consistency in anatomical phantoms by evaluating the differences between two generations of ViP phantoms, ViP1.x and ViP3.0. The RF‐induced 10g‐average electric (E‐) fields, tangential E‐fields distribution along active implantable medical devices (AIMD) routings, and estimated AIMD heating were compared for five phantoms that are part of both ViP1.x and ViP3.0. The results demonstrated that differences exceeded 3 dB (−29%, +41%) for local quantities and 1 dB (±12% for field, ±25% for power) for integrated and volume‐averaged quantities (e.g., estimated AIMD‐heating and 10 g‐average E‐fields), while the variation across different ViP phantoms of the same generation can exceed 10 dB (−68% and +217% for field, −90% and +900% for power). In conclusion, the anatomical phantom uncertainty associated with tissue assignment and segmentation quality/consistency is larger than previously assumed, i.e., 0.6 dB or ±15% (k = 1) for AIMD heating. Further, multiple phantoms based on different volunteers covering the target population are required for quantitative analysis of dosimetric endpoints, e.g., AIMD heating, which depend on patient anatomy. Phantoms with the highest fidelity in tissue assignment and segmentation should be used, as these ensure the lowest uncertainty and possible underestimation of exposure. To verify that the uncertainty decreases monotonically with improved phantom quality, the evaluation of differences between phantom generations should be repeated for any improvement in segmentation. Bioelectromagnetics. 2019;40:458–471. © 2019 Bioelectromagnetics SocietyKeywords
Funding Information
- Ph.D program of the IT'IS Foundation
This publication has 33 references indexed in Scilit:
- Exposure of the Human Body to Professional and Domestic Induction Cooktops Compared to the Basic RestrictionsBioelectromagnetics, 2012
- The Virtual Family—development of surface-based anatomical models of two adults and two children for dosimetric simulationsPhysics in Medicine & Biology, 2009
- Development of a Japanese Adult Female Voxel PhantomJournal of Nuclear Science and Technology, 2009
- Modelling the interaction of electromagnetic fields (10 MHz–10 GHz) with the human body: methods and applicationsPhysics in Medicine & Biology, 2008
- HDRK-Man: a whole-body voxel model based on high-resolution color slice images of a Korean adult male cadaverPhysics in Medicine & Biology, 2008
- Calculation of MRI‐induced heating of an implanted medical lead wire with an electric field transfer functionJournal of Magnetic Resonance Imaging, 2007
- CNMAN: a Chinese adult male voxel phantom constructed from color photographs of a visible anatomical data setRadiation Protection Dosimetry, 2007
- Japanese adult male voxel phantom constructed on the basis of CT imagesRadiation Protection Dosimetry, 2006
- Comparisons of Computed Mobile Phone Induced SAR in the SAM Phantom to That in Anatomically Correct Models of the Human HeadIEEE Transactions on Electromagnetic Compatibility, 2006
- The Visible Human ProjectProceedings of the IEEE, 1998