Abstract
Compton cameras enable medium- and high-energy gamma-ray imaging without mechanical collimation. Extending an individual detector element increases the active volume without adding readout channels, but also increases axial position uncertainty and can degrade angular resolution. This study evaluated whether a GAGG long-bar Compton camera with dual-ended depth-of-interaction (DOI) readout could mitigate this trade-off. A Geant4 detector-response model reconstructed deposited energy and axial interaction position from the two end signals. Simulated reference-bar energy and DOI resolutions were comparable in magnitude to published measurements. The proposed design was compared with a conventional two-layer planar configuration at matched crystal cross section, transverse element count, and total readout-channel count. The effects of bar length and detector gap on valid two-crystal Compton-event efficiency and the angular resolution measure (ARM) were investigated. Increasing bar length and reducing detector gap improved efficiency, although the gain diminished above approximately 75 mm and a shorter geometric baseline broadened the ARM. For an on-axis 511-keV pencil beam and the adopted event-selection criteria, a 75-mm-bar, 10-mm-gap design achieved 0.836% valid-event efficiency and an ARM full width at half maximum (FWHM) of 11.42°. The conventional configuration with the closest ARM FWHM yielded 0.172% and 12.49°, respectively, making the DOI design approximately 4.9 times as efficient at a slightly lower ARM FWHM. These results indicate that dual-ended DOI GAGG bars can improve the detector-level balance between valid-event efficiency and ARM under a fixed readout-channel budget.