Effect of Phase Encoding Direction on Image Quality in Single‐Shot EPI Diffusion‐Weighted Imaging of the Breast
- 28 February 2024
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 60 (4) , 1340-1349
Abstract
Background: In breast diffusion‐weighted imaging (DWI), distortion and physiologic artifacts affect clinical interpretation. Image quality can be optimized by addressing the effect of phase encoding (PE) direction on these artifacts.Purpose: To compare distortion artifacts in breast DWI acquired with different PE directions and polarities, and to discuss their clinical implications.Study Type: Prospective.Population: Eleven healthy volunteers (median age: 47 years old; range: 22–74 years old) and a breast phantom.Field Strength/Sequence: Single‐shot echo planar DWI and three‐dimensional fast gradient echo sequences at 3 T.Assessment: All DWI data were acquired with left–right, right–left, posterior–anterior, and anterior–posterior PE directions. In phantom data, displacement magnitude was evaluated by comparing the location of landmarks in anatomical and DWI images. Three breast radiologists (5, 17, and 23 years of experience) assessed the presence or absence of physiologic artifacts in volunteers' DWI datasets and indicated their PE‐direction preference.Statistical Tests: Analysis of variance with post‐hoc tests were used to assess differences in displacement magnitude across DWI datasets and observers. A binomial test and a chi‐squared test were used to evaluate if each in vivo DWI dataset had an equal probability (25%) of being preferred by radiologists. Inter‐reader agreement was evaluated using Gwet's AC1 agreement coefficient. A P‐value P = 0.545). In the in vivo data, there were no physiological artifacts observed in any dataset, regardless of PE direction. In the reader study, there was a significant preference for the posterior–anterior datasets which were selected 94% of the time. There was good agreement between readers (0.936).Data Conclusion: This study showed the impact of PE direction on distortion artifacts in breast DWI. In healthy volunteers, the posterior‐to‐anterior PE direction was preferred by readers.Level of Evidence: 2Technical Efficacy: Stage 1Keywords
Funding Information
- GE Healthcare
This publication has 36 references indexed in Scilit:
- Apparent Diffusion Coefficient Reproducibility Across 3 T Scanners in a Breast Diffusion PhantomJournal of Magnetic Resonance Imaging, 2022
- Correction of Artifacts Induced by B0 Inhomogeneities in Breast MRI Using Reduced‐Field‐of‐View Echo‐Planar Imaging and Enhanced Reversed Polarity Gradient MethodJournal of Magnetic Resonance Imaging, 2021
- Test–retest repeatability and reproducibility of ADC measures by breast DWI: Results from the ACRIN 6698 trialJournal of Magnetic Resonance Imaging, 2018
- Variability and bias assessment in breast ADC measurement across multiple systemsJournal of Magnetic Resonance Imaging, 2016
- Design of a breast phantom for quantitative MRIJournal of Magnetic Resonance Imaging, 2016
- Magnetic Field Inhomogeneity Effects and T 2 * Dephasing2014
- Enhanced fat suppression technique for breast imagingJournal of Magnetic Resonance Imaging, 2012
- Function biomedical informatics research network recommendations for prospective multicenter functional MRI studiesJournal of Magnetic Resonance Imaging, 2012
- Diffusion‐weighted imaging of breast cancer: Correlation of the apparent diffusion coefficient value with prognostic factorsJournal of Magnetic Resonance Imaging, 2009
- Differentiation of clinically benign and malignant breast lesions using diffusion‐weighted imagingJournal of Magnetic Resonance Imaging, 2002