Hemolytic characteristics of three suctioning systems for use with a newly developed cardiopulmonary bypass system
Open Access
- 19 August 2018
- journal article
- research article
- Published by SAGE Publications in Perfusion
- Vol. 34 (2) , 136-142
Abstract
Introduction: We have been developing a closed-circuit cardiopulmonary bypass (CPB) system (“Dihead CPB”) for application during coronary artery bypass grafting (CABG) and valve surgery. To strive for minimal hemolysis during Dihead CPB, we compared the hemolysis caused by three different suction systems and performed a clinical study with the newly applied suction system. Materials & Methods: We evaluated the hemolysis caused by roller-pump suction, the SmartSuction® Harmony® and wall suction systems with respect to suction speed and compared the systems by means of in vitro studies. A clinical study was also performed on 15 volunteers to assess hemolysis and the adequacy of the newly applied suction system with Dihead CPB. Results: Pressure inside the suction cannula was −22.5 ± 0.1 mmHg at a maximum flow of 1.5 L/min for roller-pump suction and −43.4 ± 0.1 mmHg at −150 mmHg of the set vacuum pressure of wall suction. With the SmartSuction, the pressure inside the cannula varied from −76.3 ± 1.0 to −130.3 ± 1.5 mmHg, depending on suctioning conditions. Suction speed (to suction 50 ml of blood) was fastest with the SmartSuction (69.7 ± 3.58 s) whereas, with roller suction, it was 117.3 ± 8.47 s and with wall suction 96.9 ± 7.10 s. The SmartSuction had the highest hemolysis rate (2.00 ± 0.33%) vs. 0.61 ± 0.10% for roller suction and 0.41 ± 0.11% for wall suction (p<0.001). The clinical study with wall suction showed no significant increase in plasma free hemoglobin during or after CPB compared with before surgery. Conclusions: Wall suction had less hemolysis than roller suction or the SmartSuction in the in vitro study and the clinical study with wall suction showed efficient suction speed and acceptable hemolysis, suggesting that Dihead CPB with wall suction is feasible for CABG.Keywords
Funding Information
- Shinshu University Hospital Grant (HG0502)
This publication has 11 references indexed in Scilit:
- A new extracorporeal vacuum-assisted device to optimize cardiopulmonary bypass. Comparison with the conventional systemInteractive CardioVascular and Thoracic Surgery, 2011
- The Relationships Between Air Exposure, Negative Pressure, and HemolysisASAIO Journal, 2009
- Pivot wear of a centrifugal blood pump developed for circulatory assistJournal of Artificial Organs, 2008
- Initial Experience With a Minimized Extracorporeal Bypass System: Is There a Clinical Benefit?The Annals of Thoracic Surgery, 2005
- Aortic valve replacement with the minimal extracorporeal circulation (Jostra MECC System) versus standard cardiopulmonary bypass: A randomized prospective trialThe Journal of Thoracic and Cardiovascular Surgery, 2004
- Clinical Study of Totally Roller Pumpless Cardiopulmonary Bypass SystemArtificial Organs, 1997
- Suction‐induced hemolysis at various vacuum pressures: implications for intraoperative blood salvageTransfusion, 1996
- Systemic blood activation with open and closed venous reservoirsThe Annals of Thoracic Surgery, 1995
- A Standardized System for Describing Flow/Pressure Relationships in Vascular Access DevicesASAIO Transactions, 1991
- New Glycerol Tables1Industrial & Engineering Chemistry, 1927