Calibration of an unloaded laboratory oven and the associated uncertainty of measurements

Authors:

Abstract
The aim of this research was to describe the calibration of an unloaded laboratory oven used for drying textiles samples from respiration at (37 ± 2) °C and the detailed associated uncertainty of measurements. The calibration was performed at three temperature points (35 °C, 37 °C and 40 °C) in accordance with the DKD-R 5-7:2018 guidelines. The entire calibration system including the nine calibrated Pt100 temperature probes and the multimeter that reads temperature in °C was interfaced with a PC running a LabVIEW program for data acquisition and analysis. The temperature drift, inhomogeneity, instability and the calibration uncertainty were assessed. A small temperature drift of 0.3 °C over 48 hours at 40 °C was found. The temperature inhomogeneity expressed as the maximum absolute difference between the reference location and the other locations, Max │Tref - Ti│ was found well below the acceptable accuracy of ± 2 °C. The temporal instability expressed as the maximum difference between the average and the temperature of each Pt100, Max │T̅ - Ti│ was also smaller than the acceptance limit, ± 2 °C of the textile test. The expanded uncertainty did not exceed ± 1.24 °C indicating a good oven calibration. This investigation revealed that the calibrated oven maintains good temperature drift, homogeneity and stability making it suitable for reliable textile testing.