Conduction System Pacing With DF‐1 ICD Systems Using a Medtronic 3830 Lead: A Feasibility Study
Open Access
- 23 November 2026
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
Abstract
Background: Patients with implantable cardioverter‐defibrillators (ICDs) who require ventricular pacing are at risk for pacing‐induced cardiomyopathy (PICM) from chronic right ventricular (RV) pacing. Conduction system pacing (CSP), particularly left bundle branch area pacing (LBBAP), offers more physiologic ventricular activation, but integration with ICD therapy remains challenging. The DF‐1 ICD connector standard provides an opportunity to incorporate a separate CSP lead while maintaining high‐voltage ICD therapy. We evaluated the feasibility of an alternative ICD configuration using a DF‐1 defibrillation lead paired with a Medtronic 3830 CSP lead connected to a dual‐chamber ICD generator. Methods: We conducted a retrospective case series of 17 consecutive patients who underwent this configuration at a single center between October 2022 and August 2025. For subgroup analyses, patients were stratified by baseline left ventricular ejection fraction (LVEF) as LVEF ≤ 50% ( n = 9) or LVEF > 50% ( n = 8). QRS duration, LVEF, left ventricular end‐systolic dimension (LVESD), tricuspid regurgitation (TR) severity, and device outcomes were assessed during follow‐up. Results: Median follow‐up was 15.0 months (IQR 6.2–19.2). The Medtronic 3830 lead was successfully implanted in all 17 patients, with confirmed LBBAP in 16 patients (94.1%) and deep septal pacing without definitive left bundle branch capture in one patient. No CSP lead revisions for mechanical or electrical failure were required. Among patients with paired measurements, LVEF increased from 41.6% ± 15.0% to 45.9% ± 14.1%, although the change was not statistically significant ( p = 0.169), while LVESD decreased from 4.33 ± 0.72 cm to 3.89 ± 0.79 cm ( p = 0.022). In paired subgroup analyses, patients with baseline LVEF ≤50% demonstrated an increase in LVEF from 31.9% ± 6.2% to 40.1% ± 14.3% ( p = 0.053) and a decrease in LVESD from 4.62 ± 0.65 cm to 4.11 ± 0.71 cm ( p = 0.048), whereas patients with baseline LVEF >50% demonstrated no significant change in either measure. QRS duration did not change significantly in either subgroup. Overall TR severity did not change significantly (Bhapkar p = 0.337), although 5 of 12 patients with paired imaging experienced at least a one‐grade increase in TR severity and no patient developed new severe TR. Two patients (11.8%) died during follow‐up, with neither death attributed to the device configuration. Conclusion: An ICD configuration using a DF‐1 defibrillation lead paired with a Medtronic 3830 CSP lead on a dual‐chamber generator was technically feasible, with successful 3830 lead implantation in all patients, confirmed LBBAP in 94.1%, and no CSP lead revisions for mechanical or electrical failure during intermediate‐term follow‐up. Patients with baseline LVEF ≤50% demonstrated numerical improvement in LVEF and a reduction in LVESD, while ventricular function remained stable in those with LVEF >50%. These exploratory findings require confirmation in larger prospective comparative studies.This publication has 17 references indexed in Scilit:
- Cardiac Resynchronization TherapyJAMA, 2026
- Safety and performance of a novel ICD lead for left bundle branch area pacing: Results from the ASCEND CSP trialHeart Rhythm, 2026
- Conduction System Pacing vs Right Ventricular Pacing in Preserved or Mildly Reduced Ejection Fraction: A Meta-Analysis of Randomised TrialsCanadian Journal of Cardiology, 2026
- Safety and efficacy of conduction system pacing in dual-chamber Pacemakers in a CRT-indicated population: Primary analysis of the CORE-CPP studyHeart Rhythm, 2026
- Effect of leadless left ventricular endocardial and left bundle branch area pacing on biventricular repolarization metricsHeart Rhythm, 2025
- Incidence and Prognostic Implications of Cardiac-Implantable Device-Associated Tricuspid Regurgitation: A Meta-Analysis and Meta-Regression AnalysisThe American Journal of Cardiology, 2023
- 2023 HRS/APHRS/LAHRS guideline on cardiac physiologic pacing for the avoidance and mitigation of heart failureHeart Rhythm, 2023
- Pacing-induced cardiomyopathy: A systematic review and meta-analysis of definition, prevalence, risk factors, and managementHeart Rhythm, 2022
- Incidence and predictors of right ventricular pacing-induced cardiomyopathy in patients with complete atrioventricular block and preserved left ventricular systolic functionHeart Rhythm, 2016
- Dual-Chamber Pacing or Ventricular Backup Pacing in Patients With an Implantable DefibrillatorJAMA, 2002